Low contact force and force-time integral predict early recovery and dormant conduction revealed by adenosine after pulmonary vein isolation
Open Access
- 24 January 2015
- journal article
- research article
- Published by Oxford University Press (OUP) in EP Europace
- Vol. 17 (6), 877-883
- https://doi.org/10.1093/europace/euu329
Abstract
After pulmonary vein isolation (PVI), dormant conduction (DC) is present in at least one vein in a substantial number of patients. The present study seeks to determine whether there is a relationship between poor contact forces (CF) and the presence of DC after PVI. This prospective, operator-blinded, non-randomized dual-centre trial enrolled 34 consecutive patients with paroxysmal atrial fibrillation who were candidates for PVI. Radiofrequency (RF) energy was delivered by using an irrigated-tip force-sensing ablation catheter (Tacticath®, St Jude Medical) at pre-defined target power. The operators were blinded to the CF data at all times. A total of 1476 RF applications were delivered in 743 pre-defined PV segments. For each application, the precise location of the catheter was registered and the following data were extracted from the Tacisys® unit: application duration, minimum contact force, maximum contact force, average contact force (CF), and force–time integral (FTI). Sixty minutes after PVI, spontaneous early recovery (ER) of the left atrium (LA) to PV conduction was evaluated. In the absence of ER, the presence of a DC was evaluated by using intravenous adenosine (ATP). In the 34 patients recruited (23 males; mean age: 62 ± 9 years), all PVs were successfully isolated. At the end of the 60 min waiting period, 22 patients demonstrated at least one spontaneous ER or DC under ATP. The mean CF and FTI per PV segment differed significantly among the different veins but the sites of ER and DC were evenly distributed. However, both the minimum, the first and the mean CF and FTI per PV segment were significantly lower in the PV segments presenting either ER or DC as compared with those without ER or DC (mean CF: 4.9 ± 4.8 vs. 12.2 ± 1.65 g and mean FTI: 297 ± 291 vs. 860 ± 81 g s, P < 0.001 for both). Using multivariate analysis, both the mean CF and the FTI per lesion remained significantly associated with the risk of ER or DC. Moreover, a CF < 5 g per PV segment predicted ER+ and DC+ with a sensitivity of 71% and specificity of 82%. In contrast, ER and DC were very unlikely if RF application was performed with a mean CF > 10 g (negative predictive value: 98.7%). Both a low CF and a low FTI are associated with the ER of the PVI and DC after PVI.Keywords
This publication has 12 references indexed in Scilit:
- Pulmonary Vein Isolation Using a Pace-Capture–Guided Versus an Adenosine-Guided ApproachCirculation: Arrhythmia and Electrophysiology, 2013
- Electrical Reconnection After Pulmonary Vein Isolation Is Contingent on Contact Force During Initial TreatmentCirculation: Arrhythmia and Electrophysiology, 2013
- The relationship between contact force and clinical outcome during radiofrequency catheter ablation of atrial fibrillation in the TOCCATA studyHeart Rhythm, 2012
- Adenosine plus dipyridamole: a novel strategy to enhance adenosine-induced conduction recovery after pulmonary vein isolationEP Europace, 2012
- Using omega-3 fatty acids to treat persistent atrial fibrillation: time to fish or cut bait?Heart Rhythm, 2012
- Importance of Catheter Contact Force During Irrigated Radiofrequency Ablation: Evaluation in a Porcine Ex Vivo Model Using a Force‐Sensing CatheterJournal of Cardiovascular Electrophysiology, 2010
- Catheter Ablation Versus Antiarrhythmic Drugs for Atrial FibrillationCirculation, 2008
- Novel Contact Force Sensor Incorporated in Irrigated Radiofrequency Ablation Catheter Predicts Lesion Size and Incidence of Steam Pop and ThrombusCirculation: Arrhythmia and Electrophysiology, 2008
- Reduction of AF Recurrence After Pulmonary Vein Isolation by Eliminating ATP‐Induced Transient Venous Re‐ConductionJournal of Cardiovascular Electrophysiology, 2007
- “Dormant” Pulmonary Vein Conduction Revealed by Adenosine after Ostial Radiofrequency Catheter AblationJournal of Cardiovascular Electrophysiology, 2004