Radiation Exposure to the Surgeon During Open Lumbar Microdiscectomy and Minimally Invasive Microdiscectomy
- 1 February 2011
- journal article
- research article
- Published by Ovid Technologies (Wolters Kluwer Health) in Spine
- Vol. 36 (3), 255-260
- https://doi.org/10.1097/brs.0b013e3181ceb976
Abstract
This is a prospective in vivo study comparing radiation exposure to the surgeon during 10 minimally invasive lumbar microdiscectomy cases with 10 traditional open discectomy cases as a control. Radiation exposure to the eye, chest, and hand of the operating surgeon during minimally invasive surgery (MIS) and open lumbar microdiscectomy were measured. The Occupational Exposure Guidelines were used to calculate the allowable number of cases per year from the mean values at each of the 3 sites. Fluoroscopy is a source of ionizing radiation and as such, is a potential health hazard with continued exposure during surgery. Presently, radiation exposure to the surgeon during MIS lumbar microdiscectomy is unknown. Radiation exposure to the surgeon (millirads [mR]) per case was measured by digital dosimeters placed at the level of the thyroid/eye, chest, and dominant forearm. Other data collected included operative side and level, side of the surgeon, side of the x-ray source, total fluoroscopy time, and energy output. The average radiation exposure to the surgeon during open cases was thyroid/eye 0.16 ± 0.22 mR, chest 0.21 ± 0.23 mR, and hand 0.20 ± 0.14 mR. During minimally invasive cases exposure to the thyroid/eye was 1.72 ± 1.52 mR, the chest was 3.08 ± 2.93 mR, and the hand was 4.45 ± 3.75 mR. The difference between thyroid/ eye, chest, and hand exposure during open and minimally invasive cases was statistically significant (P = 0.010, P = 0.013, and P = 0.006, respectively). Surgeons standing in an adjacent substerile room during open cases were exposed to 0.2 mR per case. MIS lumbar microdiscectomy cases expose the surgeon to significantly more radiation than open microdiscectomy. One would need to perform 1623 MIS microdiscectomies to exceed the exposure limit for whole-body radiation, 8720 cases for the lens of the eye, and 11,235 cases for the hand. Standing in a substerile room during x-ray localization in open cases is not fully protective.Keywords
This publication has 18 references indexed in Scilit:
- Radiation Exposure to the Surgeon and the Patient During KyphoplastyJournal of Spinal Disorders & Techniques, 2008
- Open versus Minimally Invasive Lumbar Microdiscectomy: Comparison of Operative Times, Length of Hospital Stay, Narcotic Use and Complicationsmin - Minimally Invasive Neurosurgery, 2008
- Radiation Safety during Spine InterventionsJournal of Vascular and Interventional Radiology, 2005
- OCCUPATIONAL EXPOSURE FROM COMMON FLUOROSCOPIC PROJECTIONS USED IN ORTHOPAEDIC SURGERYJournal of Bone and Joint Surgery, 2003
- American College of Cardiology/Society for Cardiac Angiography and Interventions Clinical Expert Consensus Document on Cardiac Catheterization Laboratory Standards: A report of the American College of Cardiology Task Force on Clinical Expert Consensus Documents endorsed by the American Heart Association and the Diagnostic and Interventional Catheterization Committee of the Council on Clinical Cardiology of the AHAJournal of the American College of Cardiology, 2001
- Ionizing radiation and cancer risk: evidence from epidemiology.Radiation Research, 1998
- Power and sample size calculations: A review and computer programControlled Clinical Trials, 1990
- A dosage response curve for the one rad range: adult risks from diagnostic radiation.American Journal of Public Health, 1979
- Ionizing irradiation and the induction of clinically significant disease in the human thyroid glandThe American Journal of Medicine, 1977
- Radiation Cancer of the SkinRadiology, 1965