Force and pressure distribution using Macintosh and GlideScope laryngoscopes in normal and difficult airways: a manikin study
- 28 September 2011
- journal article
- research article
- Published by Elsevier BV in British Journal of Anaesthesia
- Vol. 108 (1), 146-151
- https://doi.org/10.1093/bja/aer304
Abstract
The forces applied to the soft tissues of the upper airway may have a deleterious effect. This study was designed to evaluate the performance of the GlideScope compared with the Macintosh laryngoscope. Twenty anaesthetists and 20 trainees attempted tracheal intubation of a Laerdal SimMan manikin. Forces and pressure distribution applied by both laryngoscope blades onto the soft upper airway tissues were measured using film pressure transducers. The minimal force needed to achieve a successful intubation, in the same simulated scenario, was measured; additionally, we considered the visualization score achieved by using the Cormack–Lehane grades. All participants applied, on average, lower force with the GlideScope than with the Macintosh in each simulated scenario. Forces [mean (sd)] applied in the normal airway scenario [anaesthetists: Macintosh 39 (22) N and GlideScope 27 (15) N; trainees: Macintosh 45 (24) N and GlideScope 21 (15) N] were lower than forces applied in the difficult airway scenario [anaesthetists: Macintosh 95 (22) N and GlideScope 66 (20) N; trainees: Macintosh 100 (38) N and GlideScope 48 (16) N]. All the intubations using the GlideScope were successful, regardless of the scenario and previous intubation experience. The average pressure on the blades was 0.13 MPa for the Macintosh and 0.07 MPa for the GlideScope, showing a higher uniformity for the latter. The GlideScope allowed the participants to obtain a successful intubation applying a lower force. A flatter and more uniform pressure distribution, a higher successful rate, and a better glottic view were observed with the GlideScope.Keywords
Funding Information
- Department of Anaesthesia at the University Campus Bio-Medico
This publication has 24 references indexed in Scilit:
- Tracheal intubation following training with the GlideScope® compared to direct laryngoscopyAnaesthesia, 2010
- Randomized controlled trial of the Pentax AWS®, Glidescope®, and Macintosh laryngoscopes in predicted difficult intubationBritish Journal of Anaesthesia, 2009
- Expected difficult tracheal intubation: a prospective comparison of direct laryngoscopy and video laryngoscopy in 200 patientsBritish Journal of Anaesthesia, 2009
- Comparison of Intubation Success of Video Laryngoscopy Versus Direct Laryngoscopy in the Difficult Airway Using High-Fidelity SimulationSimulation in healthcare : journal of the Society for Simulation in Healthcare, 2009
- Complications of Emergency Tracheal Intubation: Immediate Airway-related Consequences: Part IIJournal of Intensive Care Medicine, 2007
- Complications of Emergency Tracheal Intubation: Hemodynamic Alterations - Part IJournal of Intensive Care Medicine, 2007
- The use of the GlideScope ® for tracheal intubation in patients with ankylosing spondylitisBritish Journal of Anaesthesia, 2006
- The GlideScope® Video Laryngoscope: randomized clinical trial in 200 patientsBritish Journal of Anaesthesia, 2005
- Tracheal intubation using a Macintosh laryngoscope or a GlideScope ® in 15 patients with cervical spine immobilizationBritish Journal of Anaesthesia, 2003
- UNEXPECTED, DIFFICULT LARYNGOSCOPY: A PROSPECTIVE SURVEY IN ROUTINE GENERAL SURGERYBritish Journal of Anaesthesia, 1991