The design and operation of a facility in which a magnetic resonance imaging (MRI) scanner is incorporated into a room used for surgical or endovascular cardiac interventions presents several challenges. MR safety must be maintained in the presence of a much wider variety of equipment than is found in a diagnostic unit, and of staff unfamiliar with the MRI environment, without compromising the safety and practicality of the interventional procedure. Both the MR‐guided cardiac interventional unit at Kings College London and the intraoperative imaging suite at the National Hospital for Neurology and Neurosurgery are single‐room interventional facilities incorporating 1.5 T cylindrical‐bore MRI scanners. The two units employ similar strategies to maintain MR safety, both in original design and day‐to‐day operational workflows, and between them over a decade of incident‐free practice has been accumulated. This article outlines these strategies, highlighting both similarities and differences between the units, as well as some lessons learned and resulting procedural changes made in both units since installation.J. Magn. Reson. Imaging 2014.This publication has 23 references indexed in Scilit:
- ACR guidance document on MR safe practices: 2013Journal of Magnetic Resonance Imaging, 2013
- AMENDMENT TO THE ICNIRP “STATEMENT ON MEDICAL MAGNETIC RESONANCE (MR) PROCEDURES: PROTECTION OF PATIENTS”Health Physics, 2009
- Origins of Intraoperative MRINeurosurgery Clinics of North America, 2009
- GUIDELINES ON LIMITS OF EXPOSURE TO STATIC MAGNETIC FIELDSHealth Physics, 2009
- INTRAOPERATIVE MAGNETIC RESONANCE IMAGING AT 3-T USING A DUAL INDEPENDENT OPERATING ROOM-MAGNETIC RESONANCE IMAGING SUITENeurosurgery, 2008
- THE NEW GENERATION POLESTAR N20 FOR CONVENTIONAL NEUROSURGICAL OPERATING ROOMSOperative Neurosurgery, 2008
- Electromagnetic field exposure limitation and the future of MRIThe British Journal of Radiology, 2005
- Endovascular interventional MRIJournal of Magnetic Resonance Imaging, 2005
- Visualization and tracking of an inflatable balloon catheter using SSFP in a flow phantom and in the heart and great vessels of patientsMagnetic Resonance in Medicine, 2004
- American College of Radiology White Paper on MR SafetyAmerican Journal of Roentgenology, 2002