Stress distribution of three NiTi rotary files under bending and torsional conditions using a mathematic analysis
- 16 December 2008
- journal article
- research article
- Published by Wiley in International Endodontic Journal
- Vol. 42 (1), 14-21
- https://doi.org/10.1111/j.1365-2591.2008.01481.x
Abstract
Aim To compare and evaluate the stress distribution of three NiTi instruments of various cross‐sectional configurations under bending or torsional condition using a finite‐element analysis model. Methodology Three NiTi files (ProFile, ProTaper and ProTaper Universal) were scanned using Micro‐CT to produce a three‐dimensional digital model. The behaviour of the instrument under bending or torsional loads was analysed mathematically in software (ABAQUS V6.5‐1), taking into consideration the nonlinear mechanical characteristic of NiTi material. Results ProFile showed the greatest flexibility, followed by ProTaper Universal and ProTaper. The highest stress was observed at the surface near the cutting edge and the base of (opposing) flutes during cantilever bending. Concentration of stresses was observed at the bottom of the flutes in ProFile and ProTaper Universal instruments in torsion. The stress was more evenly distributed over the surface of ProTaper initially, which then concentrated at the middle of the convex sides when the amount of angular deflection was increased. Conclusion Incorporating a U‐shaped groove in the middle of each side of the convex‐triangular design lowers the flexural rigidity of the origin ProTaper design. Bending leads to the highest surface stress at or near the cutting edge of the instrument. Stress concentration occurs at the bottom of the flute when the instrument is subjected to torsion.Keywords
This publication has 36 references indexed in Scilit:
- Impact of Lubricant Parameters on Rotary Instrument Torque and ForceJournal of Endodontics, 2007
- Defects in ProTaper S1 instruments after clinical use: fractographic examinationInternational Endodontic Journal, 2005
- Relative efficacy of three Ni-Ti file systems used by undergraduatesJournal of Korean Academy of Conservative Dentistry, 2005
- Shaping ability of ProTaper nickel‐titanium files in simulated resin root canalsInternational Endodontic Journal, 2004
- K3 Endo, ProTaper, and ProFile Systems: Breakage and Distortion in Severely Curved Roots of MolarsJournal of Endodontics, 2004
- Evaluation of Single-use Rotary Nickel-titanium InstrumentsJournal of Endodontics, 2003
- Comparative Analysis of Torsional and Bending Stresses in Two Mathematical Models of Nickel-Titanium Rotary Instruments: ProTaper versus ProFileJournal of Endodontics, 2003
- Effect of various canal preparation techniques using rotary nickel-titanium files on the maintenance of canal curvatureJournal of Korean Academy of Conservative Dentistry, 2003
- The efficiency of the Ni-Ti rotary files in curved simulated canals shaped by novice operatorsJournal of Korean Academy of Conservative Dentistry, 2003
- Relationship between file size and stiffness of nickel titanium instrumentsDental Traumatology, 1995