Characterization of a 2D ion chamber array for the verification of radiotherapy treatments
Open Access
- 6 July 2005
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 50 (14), 3361-3373
- https://doi.org/10.1088/0031-9155/50/14/012
Abstract
In this study we investigated the characteristics of a commercial ion chamber array and its performance in the verification of radiotherapy plans. The device was the 2D Array Seven29 model (PTW, Freiburg, Germany). This is a two-dimensional detector array with 729 ionization chambers uniformly arranged in a 27 x 27 matrix with an active area of 27 x 27 cm(2). The detector short-, medium- and long-term reproducibility have been tested through an extensive set of repeated measurements. Short-term reproducibility was well within 0.2%. Medium- and long-term reproducibility were within 1%, including set-up reproducibility errors and linac output fluctuations. Dose linearity was also assessed. The system response to dose was verified to be linear within the range 2-500 MU. Output factors matched very well pinpoint chamber measurements performed in the same experimental conditions with a maximum local percentage difference of 0.4%. Furthermore, the 2D Array sensitivity to millimetric collimator positional changes and to perturbation effect of irradiated area was tested. The comparison with ion chamber data carried out in water was very satisfying. Finally, measurements of wedge-modulated fields and IMRT beam sequence matched very well ion chamber dose profiles acquired in a water tank. The extensive tests performed in this investigation show that the 2D Array Seven29 is a reliable and accurate dosimeter and that it could be a useful tool for the quality assurance and the verification of radiotherapy plans.This publication has 23 references indexed in Scilit:
- Clinical and physical quality assurance for intensity modulated radiotherapy of prostate cancerRadiotherapy and Oncology, 2004
- Implementing IMRT in clinical practice: a joint document of the American Society for Therapeutic Radiology and Oncology and the American Association of Physicists in MedicineInternational Journal of Radiation Oncology*Biology*Physics, 2004
- Correction for dose-response variations in a scanning liquid ion chamber EPID as a function of linac gantry anglePhysics in Medicine & Biology, 2004
- Dosimetric characterization of a large area pixel-segmented ionization chamberMedical Physics, 2004
- Potential gains for irradiation of chest wall and regional nodes with intensity modulated radiotherapyInternational Journal of Radiation Oncology*Biology*Physics, 2003
- Intensity-modulated radiotherapy in nasopharyngeal carcinoma: dosimetric advantage over conventional plans and feasibility of dose escalationInternational Journal of Radiation Oncology*Biology*Physics, 2003
- A 2-D diode array and analysis software for verification of intensity modulated radiation therapy deliveryMedical Physics, 2003
- Intensity-modulated radiotherapy: current status and issues of interestInternational Journal of Radiation Oncology*Biology*Physics, 2001
- American Association of Physicists in Medicine Radiation Therapy Committee Task Group 53: Quality assurance for clinical radiotherapy treatment planningMedical Physics, 1998
- Comprehensive QA for radiation oncology: Report of AAPM Radiation Therapy Committee Task Group 40Medical Physics, 1994