Reducing Patient Radiation Exposure From CT Fluoroscopy–Guided Lumbar Spine Pain Injections by Targeting the Planning CT
- 1 February 2016
- journal article
- Published by American Roentgen Ray Society in American Journal of Roentgenology
- Vol. 206 (2), 390-394
- https://doi.org/10.2214/ajr.15.14436
Abstract
OBJECTIVE. CT fluoroscopy–guided lumbar spine pain injections typically include a preprocedural planning CT that contributes considerably to patient dose. The purpose of this study was to quantify the degree of radiation exposure reduction achieved by modifying only the planning CT component of the examination. MATERIALS AND METHODS. A retrospective review was performed of 80 CT fluoroscopy–guided lumbar spine injections. Forty patients were scanned with a standard protocol using automatic tube current modulation (method A). Another 40 patients were scanned using a new technique that fixed the tube current of the planning CT to either 50 or 100 mA on the basis of the patient's anteroposterior diameter and that reduced the z-axis coverage (method B). Dose-length products (DLPs) were compared for the two methods. RESULTS. The mean maximal tube current for the planning CT was 435.0 mA for method A and 67.5 mA for method B. The mean z-axis was shorter for method B at 6.5 cm than for method A at 9.6 cm (p < 0.0001). The mean DLP for the planning CT was 11 times lower for method B than for method A: 27.9 versus 313.1 mGy × cm, respectively (p < 0.0001). When method B was used, the mean DLP for the total procedure (i.e., planning CT plus CT fluoroscopy components) was reduced by 78%. There was no significant difference between methods A and B in CT fluoroscopy time (p = 0.37). All procedures were technically successful. CONCLUSION. A nearly fivefold reduction in radiation exposure can be achieved in CT fluoroscopy–guided lumbar spine pain injections through modifications to the planning CT alone.Keywords
This publication has 23 references indexed in Scilit:
- Radiation Dose Reduction in CT-Guided Spine Biopsies Does Not Reduce Diagnostic YieldAmerican Journal of Neuroradiology, 2014
- Radiation Dose Optimization for CT-Guided Interventional Procedures in the Abdomen and PelvisJournal of the American College of Radiology, 2014
- Radiation exposure in CT-guided interventionsEuropean Journal of Radiology, 2013
- CT-Guided Lumbar Nerve Root Injections: Are We Using the Correct Radiation Dose Settings?American Journal of Neuroradiology, 2012
- Radiation Dose Exposure for Lumbar Spine Epidural Steroid Injections: A Comparison of Conventional Fluoroscopy Data and CT Fluoroscopy TechniquesAmerican Journal of Roentgenology, 2011
- Reducing Patient Radiation Dose during CT-Guided Procedures: Demonstration in Spinal Injections for PainAmerican Journal of Neuroradiology, 2011
- Volume CT Dose Index and Dose-Length Product Displayed during CT: What Good Are They?Radiology, 2011
- Effect of Patient Size on Radiation Dose for Abdominal MDCT with Automatic Tube Current Modulation: Phantom StudyAmerican Journal of Roentgenology, 2008
- CT-Fluoroscopy: Tool or Gimmick?CardioVascular and Interventional Radiology, 2001
- Real-time CT Fluoroscopy: Evolution of an Interventional ToolRadiology, 1999