Improving the depth sensitivity of time-resolved measurements by extracting the distribution of times-of-flight
Open Access
- 15 February 2013
- journal article
- Published by Optica Publishing Group in Biomedical Optics Express
- Vol. 4 (3), 447-459
- https://doi.org/10.1364/BOE.4.000447
Abstract
Time-resolved (TR) techniques provide a means of discriminating photons based on their time-of-flight. Since early arriving photons have a lower probability of probing deeper tissue than photons with long time-of-flight, time-windowing has been suggested as a method for improving depth sensitivity. However, TR measurements also contain instrument contributions (instrument-response-function, IRF), which cause temporal broadening of the measured temporal point-spread function (TPSF) compared to the true distribution of times-of-flight (DTOF). The purpose of this study was to investigate the influence of the IRF on the depth sensitivity of TR measurements. TPSFs were acquired on homogeneous and two-layer tissue-mimicking phantoms with varying optical properties. The measured IRF and TPSFs were deconvolved using a stable algorithm to recover the DTOFs. The microscopic Beer-Lambert law was applied to the TPSFs and DTOFs to obtain depth-resolved absorption changes. In contrast to the DTOF, the latest part of the TPSF was not the most sensitive to absorption changes in the lower layer, which was confirmed by computer simulations. The improved depth sensitivity of the DTOF was illustrated in a pig model of the adult human head. Specifically, it was shown that dynamic absorption changes obtained from the late part of the DTOFs recovered from TPSFs acquired by probes positioned on the scalp were similar to absorption changes measured directly on the brain. These results collectively demonstrate that this method improves the depth sensitivity of TR measurements by removing the effects of the IRF.Keywords
This publication has 33 references indexed in Scilit:
- Variance of time-of-flight distribution is sensitive to cerebral blood flow as demonstrated by ICG bolus-tracking measurements in adult pigsBiomedical Optics Express, 2013
- Blood transfusions increase cerebral, splanchnic, and renal oxygenation in anemic preterm infantsTransfusion, 2010
- Noninvasive observation of skeletal muscle contraction using near-infrared time-resolved reflectance and diffusing-wave spectroscopyJournal of Biomedical Optics, 2010
- Changes in cerebral blood flow after acetazolamide: an experimental study comparing near‐infrared spectroscopy and SPECTEuropean Journal of Neurology, 2009
- Near infrared optical tomography using NIRFAST: Algorithm for numerical model and image reconstructionCommunications in Numerical Methods in Engineering, 2008
- Double-layer estimation of intra- and extracerebral hemoglobin concentration with a time-resolved systemJournal of Biomedical Optics, 2008
- Accuracy limits in the determination of absolute optical properties using time-resolved NIR spectroscopyMedical Physics, 2001
- Multichannel photon counting instrument for spatially resolved near infrared spectroscopyReview of Scientific Instruments, 1999
- Improved solutions of the steady-state and the time-resolved diffusion equations for reflectance from a semi-infinite turbid mediumJournal of the Optical Society of America A, 1997
- Time resolved reflectance and transmittance for the noninvasive measurement of tissue optical propertiesApplied Optics, 1989