Tilt and decentration of spherical and aspheric intraocular lenses: Effect on higher-order aberrations
- 1 June 2009
- journal article
- research article
- Published by Ovid Technologies (Wolters Kluwer Health) in Journal of Cataract & Refractive Surgery
- Vol. 35 (6), 1006-1012
- https://doi.org/10.1016/j.jcrs.2009.01.023
Abstract
PURPOSE: To determine the impact of spherical and aspheric foldable intraocular lens (IOL) tilt and decentration on optical quality after cataract surgery in an intraindividual comparative study. SETTING: Department of Ophthalmology, Goethe-University, Frankfurt am Main, Germany. METHODS: Randomized implantation of a spherical IOL (Sensar AR40e) was performed in 1 eye and of an aspheric IOL with negative spherical aberration (Tecnis Z9000) in the contralateral eye. Three to 4 months postoperatively, the wavefront was measured and higher-order aberrations (HOAs) were calculated for virtual pupil diameters of 3.5 mm and 6.0 mm. Tilt and decentration of the IOLs were measured using Scheimpflug photography. The tilt and decentration, HOAs, and best corrected visual Strehl ratio of the optical transfer function (BCVSOTF) calculated from the wavefront aberration were compared. The effect of tilt and decentration on HOAs and optical quality was assessed using multiple regression analysis. RESULTS: The mean optic tilt was 2.89 degrees ± 1.46 (SD) for the spherical IOL and 2.85 ± 1.36 degrees for the aspheric IOL. The mean optic decentration was 0.19 ± 0.12 mm and 0.27 ± 0.16 mm, respectively. No significant intergroup differences in IOL tilt or decentration were found. Tilt and decentration did not significantly affect the BCVSOTF with either IOL. CONCLUSIONS: The amounts of tilt and decentration of both IOLs were not large enough to cause deterioration of optical quality. Thus, when IOL tilt and decentration are within normal limits, they do not compromise the correction of spherical aberration by the aspheric IOL.Keywords
This publication has 37 references indexed in Scilit:
- Tilt and decentration of intraocular lenses in vivo from Purkinje and Scheimpflug imagingJournal of Cataract & Refractive Surgery, 2007
- Intraindividual comparison of higher-order aberrations after implantation of aspherical and spherical intraocular lenses as a function of pupil diameterJournal of Cataract & Refractive Surgery, 2006
- Tilt and Decentration of Three-Piece Foldable High-Refractive Silicone and Hydrophobic Acrylic Intraocular Lenses With 6-mm Optics in an Intraindividual ComparisonAmerican Journal of Ophthalmology, 2005
- Effect of tilt of 2 acrylic intraocular lenses on high-order aberrationsJournal of Cataract & Refractive Surgery, 2005
- Influence of intraocular lens optical design on high-order aberrationsJournal of Cataract & Refractive Surgery, 2005
- Optical performance of 3 intraocular lens designs in the presence of decentrationJournal of Cataract & Refractive Surgery, 2005
- Influence of intraocular lens tilt and decentration on wavefront aberrationsJournal of Cataract & Refractive Surgery, 2004
- Improved functional vision with a modified prolate intraocular lensJournal of Cataract & Refractive Surgery, 2004
- Zur Entwicklungsgeschichte der IntraokularlinsenDer Ophthalmologe, 2001
- Scheimpflug Records without Distortion – A Mythos?Ophthalmic Research, 1999