CBF Measured by Xe-CT: Approach to Analysis and Normal Values

Abstract
Normal reference values and a practical approach to CBF analysis are needed for routine clinical analysis and interpretation of xenon-enhanced computed tomography (CT) CBF studies. We measured CBF in 67 normal individuals with the GE 9800 CT scanner adapted for CBF imaging with stable Xe. CBF values for vascular territories were systematically analyzed using the clustering of contiguous 2-cm circular regions of interest (ROIs) placed within the cortical mantle and basal ganglia. Mixed cortical flows averaged 51 ± 10 ml · 100 g−1 · min−1. High and low flow compartments, sampled by placing 5-mm circular ROIs in regions containing the highest and lowest flow values in each hemisphere, averaged 84 ± 14 and 20 ± 5 ml · 100 g−1 · min−1, respectively. Mixed cortical flow values as well as values within the high flow compartment demonstrated significant decline with age; however, there were no significant age-related changes in the low flow compartment. The clustering of systematically placed cortical and subcortical ROIs has provided a normative data base for Xe-CT CBF and a flexible and uncomplicated method for the analysis of CBF maps generated by Xe-enhanced CT.