An ultra-low field (ULF) magnetic resonance imaging (MRI) system was set up in an urban laboratory without magnetic shielding. The measured environmental gradient fields of 1 ∼ 5 μT/m caused image distortion. We designed a gradient detection and compensation system to effectively balance the gradient tensor components. The free induction decay signal duration of tap water was thus extended from 0.3 s to 2.5 s, providing the possibility for high-resolution imaging. Two-dimensional MRI images were then obtained at 130 μT with a helium-cooled second-order superconducting quantum interference device gradiometer. This result allows us to develop an inexpensive ULF MRI system for biological studies.Keywords
Funding Information
- National Natural Science Foundation of China (11204339)
This publication has 11 references indexed in Scilit:
- Low-field MRI measurements using a tuned HTS SQUID as detector and permanent magnet pre-polarization fieldSuperconductor Science and Technology, 2012
- Magnetic Field Improved ULF-NMR Measurement in an Unshielded Laboratory Using a Low-Tc SQUIDPhysics Procedia, 2012
- Microtesla MRI of the human brain combined with MEGJournal of Magnetic Resonance, 2008
- Detection of proton NMR signal in the Earth’s magnetic field at an urban laboratory environment without shieldingSuperconductor Science and Technology, 2008
- Magnetic full-tensor SQUID gradiometer system for geophysical applicationsGeophysics, 2006
- SQUID-detected microtesla MRI in the presence of metalJournal of Magnetic Resonance, 2005
- SQUID‐detected MRI at 132 μT with T1‐weighted contrast established at 10 μT–300 mTMagnetic Resonance in Medicine, 2004
- GETMAG – a SQUID Magnetic Tensor Gradiometer for Mineral and Oil ExplorationExploration Geophysics, 2004
- Planar HTS gradiometers with large baselineIEEE Transactions on Applied Superconductivity, 1997
- Room temperature three sensor magnetic field gradiometerReview of Scientific Instruments, 1996