Specific-heat measurement of single metallic, carbon, and ceramic fibers at very high temperature
- 1 June 2005
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 76 (6)
- https://doi.org/10.1063/1.1927102
Abstract
The main objective of this work is to present a method for measuring the specific heat of single metallic, carbon, and ceramic fibers at very high temperature. The difficulty of the measurement is due to the microscale of the fiber (approximate to 10 mu m) and the important range of temperature (700-2700 K). An experimental device, a modelization of the thermal behavior, and an analytic model have been developed. A discussion on the measurement accuracy yields a global uncertainty lower than 10%. The characterization of a tungsten filament with thermal properties identical to those of the bulk allows the validation of the device and the thermal estimation method. Finally, measurements on carbon and ceramic fibers have been done at very high temperature. (c) 2005 American Institute of Physics.This publication has 5 references indexed in Scilit:
- The tensile behavior of carbon fibers at high temperatures up to 2400 °CCarbon, 2004
- Heat capacity of oxide glasses measured by AC calorimetryJournal of Non-Crystalline Solids, 2002
- Electrical AC behaviour of MPCVD diamond Schottky diodesDiamond and Related Materials, 2001
- Introduction to CeramicsJournal of the Electrochemical Society, 1977
- Thermal conductivity, electric resistivity and specific heat of dense graphitesCarbon, 1970