Comparison of four energy-based vascular sealing and cutting instruments: A porcine model
- 20 December 2007
- journal article
- research article
- Published by Springer Science and Business Media LLC in Surgical Endoscopy
- Vol. 22 (2), 534-538
- https://doi.org/10.1007/s00464-007-9619-8
Abstract
To compare the safety and efficacy of four energy-based vascular sealing and cutting instruments. Blood vessels of various types and diameters were harvested from four pigs using four instruments: Harmonic ACE™ (Ethicon Endo-Surgery, Cincinnati, OH), LigaSure™ V and LigaSure Atlas™ (Valleylab, Inc., Boulder, CO; a division of Tyco Healthcare), and EnSeal™ vessel fusion system (SurgRx, Inc. Redwood City, CA). The diameters of the vessels, speed and adequacy of the cutting and sealing process, and bursting pressures were compared. An additional set of specimens was sealed and left in situ for up to 4 h after which the vessels were harvested and histopathologically analyzed for the degree of thermal injury. The bursting pressures were significantly higher with EnSeal™ compared to all other instruments (p < 0.0001). The sealing process was significantly shorter with Harmonic ACE™ and significantly longer with LigaSure Atlas™ (p <0.0001). The mean seal width was larger with the LigaSure Atlas™ compared to the other instruments, and it was smaller with EnSeal™ and Harmonic ACE™. Less radial adventitial collagen denaturation was present with EnSeal™ and LigaSure™ V than with the other two instruments; there were no significant differences in collagen denaturation although proximal thermal injury to the smooth muscle in the media of the vessel wall was less common with LigaSure Atlas™ than with the other instruments; however, the numbers were too small for statistical analysis. The bursting pressures with EnSeal™ were significantly higher than with all the other instruments. Harmonic ACE™ was the fastest sealing instrument and LigaSure Atlas™ was slowest. EnSeal™ created less radial thermal damage to the adventitial collagen of the vessels and LigaSure Atlas™ created less thermal damage to the media of the vessels. The clinical significance of these findings is unknown.Keywords
This publication has 11 references indexed in Scilit:
- Comparison of Monopolar Electrocoagulation, Bipolar Electrocoagulation, Ultracision, and LigasureSurgery Today, 2006
- Efficacy and quality of vessel sealingSurgical Endoscopy, 2006
- Vascular pedicle ligation techniques during laparoscopic colectomySurgical Endoscopy, 2005
- Energy Sources for Laparoscopic Colectomy: A Prospective Randomized Comparison of Conventional Electrosurgery, Bipolar Computer-Controlled Electrosurgery and Ultrasonic Dissection. Operative Outcome and Costs AnalysisSurgical Innovation, 2005
- Bursting Tension of Medium and Large Canine Arteries Sealed with Ultrasonic Energy or Suture LigationVeterinary Surgery, 2005
- A Comparison of Laparoscopic Bipolar Vessel Sealing Devices in the Hemostasis of Small-, Medium-, and Large-Sized ArteriesJournal of Laparoendoscopic & Advanced Surgical Techniques, 2003
- Comparison of ultrasonic energy, bipolar thermal energy, and vascular clips for the hemostasis of small-, medium-, and large-sized arteriesSurgical Endoscopy, 2003
- Evaluation of a vessel sealing system, bipolar electrosurgery, harmonic scalpel, titanium clips, endoscopic gastrointestinal anastomosis vascular staples and sutures for arterial and venous ligation in a porcine modelPublished by Rockefeller University Press ,2003
- Evaluation of a Vessel Sealing System, Bipolar Electrosurgery, Harmonic Scalpel, Titanium Clips, Endoscopic Gastrointestinal Anastomosis Vascular Staples and Sutures for Arterial and Venous Ligation in a Porcine ModelJournal of Urology, 2003
- High-burst-strength, feedback-controlled bipolar vessel sealingSurgical Endoscopy, 1998