Measurement of functional residual capacity by modified multiple breath nitrogen washout for spontaneously breathing and mechanically ventilated patients
Open Access
- 12 July 2011
- journal article
- research article
- Published by Elsevier BV in British Journal of Anaesthesia
- Vol. 107 (5), 796-805
- https://doi.org/10.1093/bja/aer220
Abstract
There is a need for a bedside functional residual capacity (FRC) measurement method that performs well in intensive care patients during many modes of ventilation including controlled, assisted, spontaneous, and mixed. We developed a modified multiple breath nitrogen washout method for FRC measurement that relies on end-tidal gas fractions and alveolar tidal volume measurements as inputs but does not require the traditional measurements of volume of nitrogen or oxygen. Using end-tidal measurements, not volume, reduces errors from signal synchronization. This study was designed to assess the accuracy, precision, and repeatability of the proposed FRC system in subjects with variable ventilation patterns including some spontaneous effort. The accuracy and precision of measurements were assessed by comparing the novel N2 washout FRC values to the gold standard, body plethysmography, in 20 spontaneously breathing volunteers. Repeatability was assessed by comparing subsequent measurements in 20 intensive care patients whose lungs were under controlled and assisted mechanical ventilation. Compared with body plethysmography, the accuracy (mean bias) of the novel method was −0.004 litre and precision [1 standard deviation (sd)] was 0.209 litre [mean (sd)] [−0.1 (5.9)% of body plethysmography]. The difference between repeated measurements was 0.009 (0.15) litre [mean (sd)] [0.4 (6.4)%]. The coefficient of repeatability was 0.31 litre (12.7%). The modified multiple breath nitrogen washout method for FRC measurement provides improved precision and equivalent accuracy and repeatability compared with existing methods during ventilation with variable ventilation patterns. Further study of the novel N2 washout method is needed.Keywords
This publication has 27 references indexed in Scilit:
- The Impact of Different Step Changes of Inspiratory Fraction of Oxygen on Functional Residual Capacity Measurements Using the Oxygen Washout Technique in Ventilated PatientsAnesthesia & Analgesia, 2008
- Measurement of functional residual capacity by helium dilution during partial support ventilation: in vitro accuracy and in vivo precision of the methodIntensive Care Medicine, 2007
- Determination of functional residual capacity by oxygen washin-washout: a validation studyIntensive Care Medicine, 2007
- The Accuracy of the Oxygen Washout Technique for Functional Residual Capacity Assessment During Spontaneous BreathingAnesthesia & Analgesia, 2007
- Estimation of Functional Residual Capacity at the Bedside Using Standard Monitoring Equipment: A Modified Nitrogen Washout/Washin Technique Requiring a Small Change of the Inspired Oxygen FractionAnesthesia & Analgesia, 2005
- Functional Residual Capacity and Respiratory Mechanics as Indicators of Aeration and Collapse in Experimental Lung InjuryAnesthesia & Analgesia, 2004
- Lung volume in mechanically ventilated patients: measurement by simplified helium dilution compared to quantitative CT scanIntensive Care Medicine, 2004
- The recording of FRC — is it of importance and can it be made simple?Intensive Care Medicine, 1993
- Automated sulfur hexafluoride washout functional residual capacity measurement system for any mode of mechanical ventilation as well as spontaneous respirationCritical Care Medicine, 1990
- A RAPID PLETHYSMOGRAPHIC METHOD FOR MEASURING THORACIC GAS VOLUME: A COMPARISON WITH A NITROGEN WASHOUT METHOD FOR MEASURING FUNCTIONAL RESIDUAL CAPACITY IN NORMAL SUBJECTS 1JCI Insight, 1956