Abstract
Commissioning data have been measured for an Elekta Precise linear accelerator running at 6 MV without a flattening filter with the aim of studying the effects of flattening filter removal on machine operation and beam characterization. Modern radiotherapy practice now routinely relies on the use of fluence modifying techniques such as IMRT, i.e. the active production of non-flat beams. For these techniques the flattening filter should not be necessary. It is also possible that the increased intensity around the central axis associated with unflattened beams may be useful for conventional treatment planning by acting as a field-in-field or integrated boost technique. For this reason open and wedged field data are presented. Whilst problems exist in running the machine filter free clinically, this paper shows that in many ways the beam is actually more stable, exhibiting almost half the variation in field symmetry for changes in steering and bending currents. Dosimetric benefits are reported here which include a reduction in head scatter by approx. 70%, decreased penumbra (0.5 mm), lower dose outside of the field edge (11%) and a doubling in dose rate (2.3 times for open and 1.9 times for wedged fields). Measurements also show that reduced scatter also reduces leakage radiation by approx. 60%, significantly lowering whole body doses. The greatest benefit of filter-free use is perceived to be for IMRT where increased dose rate combined with reduced head scatter and leakage radiation should lead to improved dose calculation, giving simpler, faster and more accurate dose delivery with reduced dose to normal tissues.