Assessment of left ventricular mass and volumes by three-dimensional echocardiography in patients with or without wall motion abnormalities: comparison against cine magnetic resonance imaging

Abstract
Aim: To evaluate if three-dimensional echocardiography (3-DE) is as accurate and reproducible as cine magnetic resonance imaging (cMR) in estimating left ventricular (LV) parameters in patients with and without wall motion abnormalities (WMA). Methods: 83 patients (33 with WMA) underwent 3-DE and cMR. 3-DE datasets were analysed using a semi-automatic contour detection algorithm. The accuracy of 3-DE was tested against cMR in the two groups of patients. All measurements were made twice by two different observers. Results: LV mass by 3-DE was similar to that obtained by cMR (149 (SD 42) g vs 148 (45) g, p = 0.67), with small bias (1 (28) g) and excellent interobserver agreement (−2 (31) g vs 4 (26) g). The two measurements were also highly correlated (r = 0.94), irrespective of WMA. End-diastolic and end-systolic LV volumes and ejection fraction by 3-DE and cMR were highly correlated (r = 0.97, 0.98, 0.94, respectively). Yet, 3-DE underestimated cMR end-diastolic volumes (167 (68) ml vs 187 (70) ml, pConclusion: 3-DE permits accurate determination of LV mass and volumes irrespective of the presence or absence of WMA. LV parameters obtained by 3-DE are also as reproducible as those obtained by cMR. This suggests that 3-DE can be used to follow up patients with LV hypertrophy and/or remodelling.

This publication has 30 references indexed in Scilit: