Anemia causes hypoglycemia in intensive care unit patients due to error in single-channel glucometers: Methods of reducing patient risk*
- 1 February 2010
- journal article
- research article
- Published by Ovid Technologies (Wolters Kluwer Health) in Critical Care Medicine
- Vol. 38 (2), 471-476
- https://doi.org/10.1097/ccm.0b013e3181bc826f
Abstract
Intensive insulin therapy in the critically ill reduces mortality but carries the risk of increased hypoglycemia. Point-of-care blood glucose analysis is standard; however, anemia causes falsely high values and potentially masks hypoglycemia. Permissive anemia is practiced routinely in most intensive care units. We hypothesized that point-of-care glucometer error due to anemia is prevalent, can be corrected mathematically, and correction uncovers occult hypoglycemia during intensive insulin therapy. The study has both retrospective and prospective phases. We reviewed data to verify the presence of systematic error, determine the source of error, and establish the prevalence of anemia. We confirmed our findings by reproducing the error in an in vitro model. Prospective data were used to develop a correction formula validated by the Monte Carlo method. Correction was implemented in a burn intensive care unit and results were evaluated after 9 mos. Burn and trauma intensive care units at a single research institution. Samples for in vitro studies were taken from healthy volunteers. Samples for formula development were from critically ill patients who received intensive insulin therapy. Insulin doses were calculated based on predicted serum glucose values from corrected point-of-care glucometer measurements. Time-matched point-of-care glucose, laboratory glucose, and hematocrit values. We previously found that anemia (hematocrit <34%) produces systematic error in glucometer measurements. The error was correctible with a mathematical formula developed and validated, using prospectively collected data. Error of uncorrected point-of-care glucose ranged from 19% to 29% (p < .001), improving to ≤5% after mathematical correction of prospective data. Comparison of data pairs before and after correction formula implementation demonstrated a 78% decrease in the prevalence of hypoglycemia in critically ill and anemic patients treated with insulin and tight glucose control (p < .001). A mathematical formula that corrects erroneous point-of-care glucose values due to anemia in intensive care unit patients reduces the prevalence of hypoglycemia during intensive insulin therapy.Keywords
This publication has 32 references indexed in Scilit:
- Moving beyond tight glucose control to safe effective glucose controlCritical Care, 2008
- Error Rates Resulting From Anemia can be Corrected in Multiple Commonly Used Point-of-Care GlucometersThe Journal of Trauma and Acute Care Surgery, 2008
- Hypoglycemia in the intensive care unitCurrent Opinion in Clinical Nutrition and Metabolic Care, 2007
- Current controversies around tight glucose control in critically ill patientsCurrent Opinion in Clinical Nutrition and Metabolic Care, 2007
- Intensive Insulin Therapy in the Medical ICUThe New England Journal of Medicine, 2006
- Revisiting transfusion practices in critically ill patients*Critical Care Medicine, 2005
- Pentamidine-Induced Derangements of Glucose Homeostasis: Determinant roles of renal failure and drug accumulation: A study of 128 patientsDiabetes Care, 1995
- Reflectometric Blood Glucose Determination in the Neonatological Intensive Care: Haematocrit DependenceExperimental and Clinical Endocrinology & Diabetes, 1984
- The Monte Carlo MethodJournal of the American Statistical Association, 1949
- THE ANEMIA OF THERMAL BURNSAnnals of Surgery, 1946