Continuous Renal Replacement Therapy in Critically Ill Patients: Monitoring Circuit Function
Open Access
- 1 August 1996
- journal article
- research article
- Published by SAGE Publications in Anaesthesia and Intensive Care
- Vol. 24 (4), 423-429
- https://doi.org/10.1177/0310057x9602400402
Abstract
There is currently no universally accepted method to monitor circuit function or guidelines for circuit replacement during continuous renal replacement therapies (CRRT). The objectives of this study were to diagnose the causes of circuit failure, identify factors responsible for circuit clotting and determine a predictive monitor of circuit function. The CRRT technique used in this study was continuous venovenous haemodialysis (CVVHD). Continuous monitoring of circuit pressures (pre- and post-haemofilter and their difference: the transfilter pressure gradient) was used to diagnose the causes of circuit failure. In circuits ceasing due to clotting, the factors thought to contribute, anticoagulation, haematocrit and platelet count, were measured at the commencement of CVVHD and every eight hours thereafter until circuit failure. Monitors of circuit function, creatinine clearance and plasma to diafiltrate urea ratio were measured every eight hours and compared to the transfilter pressure gradient. During a three-month period data was collected on five consecutive patients (41 consecutive haemofilters). Clotting of the haemofilter (63%) and air detection chamber (7.5%) were the most common identifiable causes of circuit failure. The duration of their circuit life was described using multiple regression analysis, i.e. hours of filter life = -82.8 + (Δ platelet count x 0.25) + (Δ haematocrit x 3.6) + (circuit flow [ml/min] x 4) R2 = 0.77. A rise in transfilter pressure gradient and a fall in haemofilter function discriminated clotted filters with falling function (decrease in creatinine clearance and urea ratio) from unclotted filters. In any circuit an increase of 26 mmHg or more in the transfilter pressure gradient accurately predicted circuit failure due to clotting and imminent cessation of function. Increases in platelet count, haematocrit, and low circuit flows are important determinants of haemofilter life. The measurement of transfilter pressure gradient across the haemofilter is an accurate bedside monitor of circuit function.Keywords
This publication has 11 references indexed in Scilit:
- Improved cardiovascular stability during continuous modes of renal replacement therapy in critically ill patients with acute hepatic and renal failureCritical Care Medicine, 1993
- Continuous Venovenous Hemofiltration: An Alternative to Continuous Arteriovenous Hemofiltration and Hemodiafiltration in Acute Renal FailureAmerican Journal of Kidney Diseases, 1991
- Comparison of pump-driven and spontaneous continuous haemofiltration in postoperative acute renal failureThe Lancet, 1991
- Continuous arteriovenous haemodiafiltration: optimal therapy for acute renal failure in an intensive care setting?Australian and New Zealand Journal of Medicine, 1990
- Acute renal replacement – which treatment is best?Australian and New Zealand Journal of Medicine, 1990
- Continuous high volume venous-venous haemofiltration in acute renal failureIntensive Care Medicine, 1989
- Hypoxemia during hemodialysisCritical Care Medicine, 1979
- The Morphology of “Acute Tubular Necrosis” in ManMedicine, 1979
- Hemodynamic Changes during HemodialysisJMIR AI, 1978
- Arteriovenous haemofiltration: A new and simple method for treatment of over-hydrated patients resistant to diureticsKlinische Wochenschrift, 1977