Intrathoracic Impedance Monitoring in Patients With Heart Failure

Abstract
Background— Patients with heart failure are frequently hospitalized for fluid overload. A reliable method for chronic monitoring of fluid status is therefore desirable. We evaluated an implantable system capable of measuring intrathoracic impedance to identify potential fluid overload before heart failure hospitalization and to determine the correlation between intrathoracic impedance and standard measures of fluid status during hospitalization. Methods and Results— Thirty-three patients with NYHA class III and IV heart failure were implanted with a special pacemaker in the left pectoral region and a defibrillation lead in the right ventricle. Intrathoracic impedance was regularly measured and recorded between the lead and the pacemaker case. During hospitalizations, pulmonary capillary wedge pressure and fluid status were monitored. Ten patients were hospitalized for fluid overload 25 times over 20.7±8.4 months. Intrathoracic impedance decreased before each admission by an average of 12.3±5.3% ( P P r =−0.61, P r =−0.70, P Conclusions— Intrathoracic impedance is inversely correlated with pulmonary capillary wedge pressure and fluid balance and decreased before the onset of patient symptoms and before hospital admission for fluid overload. Regular monitoring of impedance may provide early warning of impending decompensation and diagnostic information for titration of medication.