Comparison of the performance between portal dosimetry and a commercial two-dimensional array system on pretreatment quality assurance for volumetric-modulated arc and intensity-modulated radiation therapy
- 1 April 2014
- journal article
- Published by Korean Physical Society in Journal of the Korean Physical Society
- Vol. 64 (8), 1207-1212
- https://doi.org/10.3938/jkps.64.1207
Abstract
The aim of this study was to compare the dosimetric performance and to evaluate the pretreatment quality assurance (QA) of a portal dosimetry and a commercial two-dimensional (2-D) array system. In the characteristics comparison study, the measured values for the dose linearity, dose rate response, reproducibility, and field size dependence for 6-MV photon beams were analyzed for both detector systems. To perform the qualitative evaluations of the 10 IMRT and the 10 VMAT plans, we used the Gamma index for quantifying the agreement between calculations and measurements. The performance estimates for both systems show that overall, minimal differences in the dosimetric characteristics exist between the Electron portal imaging device (EPID) and 2-D array system. In the qualitative analysis for pretreatment quality assurance, the EPID and 2-D array system yield similar passing rate results for the majority of clinical Intensity-modulated radiation therapy (IMRT) and Volumetric-modulated arc therapy (VMAT) cases. These results were satisfactory for IMRT and VMAT fields and were within the acceptable criteria of γ%≤1, γ avg <0.5. The EPDI and the 2-D array systems showed comparable dosimetric results. In this study, the results revealed both systems to be suitable for patient-specific QA measurements for IMRT and VMAT. We conclude that, depending on the status of clinic, both systems can be used interchangeably for routine pretreatment QA.Keywords
This publication has 11 references indexed in Scilit:
- Statistical analysis of IMRT dosimetry quality assurance measurements for local delivery guidelineRadiation Oncology, 2011
- A literature review of electronic portal imaging for radiotherapy dosimetryRadiotherapy and Oncology, 2008
- Establishing action levels for EPID‐based QA for IMRTJournal of Applied Clinical Medical Physics, 2008
- Pretreatment verification of IMRT absolute dose distributions using a commercial a-Si EPIDMedical Physics, 2006
- Clinical experience with EPID dosimetry for prostate IMRT pre‐treatment dose verificationMedical Physics, 2006
- Improving IMRT quality control efficiency using an amorphous silicon electronic portal imagerMedical Physics, 2005
- The use of an aSi-based EPID for routine absolute dosimetric pre-treatment verification of dynamic IMRT fieldsRadiotherapy and Oncology, 2004
- Dosimetric IMRT verification with a flat‐panel EPIDMedical Physics, 2003
- Dosimetric properties of an amorphous silicon electronic portal imaging device for verification of dynamic intensity modulated radiation therapyMedical Physics, 2003
- Physical aspects of a liquid-filled ionization chamber with pulsed polarizing voltageMedical Physics, 1991