Custom-designed femoral prostheses in total hip arthroplasty done with cement for severe dysplasia of the hip.
- 1 October 1993
- journal article
- Published by Ovid Technologies (Wolters Kluwer Health) in The Journal of Bone & Joint Surgery
- Vol. 75 (10), 1497-1504
- https://doi.org/10.2106/00004623-199310000-00010
Abstract
A custom-designed femoral prosthesis was implanted with cement and a standard acetabular component was used to treat nineteen severely dysplastic hips in fourteen consecutively managed patients. Components that had been custom-designed with the use of plain radiography were used because the anatomical reconstructive goals could not be achieved with commercially available implants. These goals were to match the offset of the femoral head and the length of the lower limb with those on the normal side for patients who had unilateral involvement and to provide an average (thirty to forty-millimeter) offset with equal limb lengths for patients who had bilateral involvement. A retrospective clinical and radiographic analysis was performed. The diagnoses included coxa vara (one hip), congenital dislocation (twelve hips), achondroplasia (three hips), and spondyloepiphyseal dysplasia (three hips).The mean age at the time of the reconstruction was forty-nine years (range, twenty-two to seventy-three years), and the mean duration of follow-up was fifty-seven months (range, twenty-seven to 108 months). In five hips, bone-grafting of the acetabulum was needed to obtain superolateral coverage. The clinical result was excellent in eighteen hips and good in one.No revisions have been performed to date. Two femoral components were possibly loose radiographically. One was associated with a definite loosening of the acetabular cup. In addition, one other cup was possibly loose. There was a 100 per cent rate of survival if only a revision procedure was considered as a failure. The probability of survival of the custom-designed femoral component was 93 per cent (confidence interval, 95 per cent (range, 57.5 to 99.2 per cent)) at nine years if possible radiographic failures were included. The custom implants were designed with a varus neck to reduce impingement, and there was a 53 per cent average gain in the effective abductor lever arm.The use of a custom femoral component in hips with severe dysplasia allows for a good fit of the stem, preservation of bone stock in the superomedial aspect of the proximal part of the femur, restoration of length of the lower limb, achievement of the desired offset, and, therefore, improvement of the biomechanics of the hip. The use of custom prostheses in patients in whom standard devices cannot provide anatomical restoration is justified on the basis of clinical results and apparent cost-effectiveness.Keywords
This publication has 26 references indexed in Scilit:
- Long-term results of total hip arthroplasty in congenital dislocation and dysplasia of the hip. A follow-up note.The Journal of Bone & Joint Surgery, 1991
- A New System to Produce Intraoperatively Custom Femoral Prosthesis From Measurements Taken During the Surgical ProcedureClinical Orthopaedics and Related Research, 1989
- Shape the Implant to the PatientClinical Orthopaedics and Related Research, 1989
- Fit the Patient to the ProsthesisClinical Orthopaedics and Related Research, 1989
- Total Hip Acetabular Component Position Affects Component Loosening RatesClinical Orthopaedics and Related Research, 1988
- Further follow-up on socket fixation using a metal-backed acetabular component for total hip replacement. A minimum ten-year follow-up study.The Journal of Bone & Joint Surgery, 1987
- Loosening of the femoral component after use of the medullary-plug cementing technique. Follow-up note with a minimum five-year follow-up.The Journal of Bone & Joint Surgery, 1986
- The effect of conformity, thickness, and material on stresses in ultra-high molecular weight components for total joint replacement.The Journal of Bone & Joint Surgery, 1986
- Complex total hip replacement for dysplastic or hypoplastic hips using miniature or microminiature components.The Journal of Bone & Joint Surgery, 1983
- Low-Friction Arthroplasty in Congenital Subluxation of the HipClinical Orthopaedics and Related Research, 1973