A clinical assessment of the performance of a sensor to detect crystalline biofilm formation on indwelling bladder catheters
Open Access
- 7 November 2006
- journal article
- research article
- Published by Wiley in BJU International
- Vol. 98 (6), 1244-1249
- https://doi.org/10.1111/j.1464-410x.2006.06562.x
Abstract
To test the ability of a sensor developed to signal infection by the organisms that generate the crystalline biofilms that encrust catheters, to give an early warning that encrustation was occurring on patients' catheters, as the care of many patients undergoing long-term bladder catheterization is complicated by the encrustation and blockage of their catheters. Twenty patients were followed prospectively for the lifetime of one of their catheters. Sensors based on cellulose acetate/bromothymol blue were placed in the urine-collection bags, which were changed as usual at weekly intervals. The bacteriology was assessed and pH determined weekly on urine samples. Photographic records were made of the sensors twice weekly. On removal, each catheter was examined for encrustation and blockage. Proteus mirabilis was not isolated from five patients and in these cases the sensor colour remained golden-yellow to brown. The catheters drained for the scheduled period and showed no signs of encrustation. By contrast, the sensors turned dark blue/black in the urine of all 15 patients infected with P. mirabilis. All these patients' catheters were encrusted and in 12 the catheters blocked. The mean interval between the sensor signalling and the catheter blocking was 12 days. The cellulose acetate/bromothymol blue sensors placed in the urine collection bags are capable of signalling infection by P. mirabilis. They also signal the early stages of catheter encrustation and allow catheter replacement in ample time to avoid the clinical crises and emergency referrals caused by catheter blockage.Keywords
This publication has 14 references indexed in Scilit:
- A Sensor To Detect the Early Stages in the Development of CrystallineProteus mirabilisBiofilm on Indwelling Bladder CathetersJournal of Clinical Microbiology, 2006
- Effect of triclosan on the development of bacterial biofilms by urinary tract pathogens on urinary cathetersJournal of Antimicrobial Chemotherapy, 2005
- Prospective study of individuals with long‐term urinary catheters colonized with Proteus speciesBJU International, 2005
- A Strategy for the Control of Catheter Blockage by Crystalline Proteus mirabilis Biofilm Using the Antibacterial Agent TriclosanEuropean Urology, 2005
- Factors modulating the pH at which calcium and magnesium phosphates precipitate from human urineUrological Research, 2005
- Long‐term catheterization of the bladder:BJU International, 1996
- Complications of urinary tract infections associated with devices used for long-term bladder managementJournal of Hospital Infection, 1994
- The characteristics and management of patients with recurrent blockage of long‐term urinary cathetersJournal of Advanced Nursing, 1994
- Infection of catheterised patients: bacterial colonisation of encrusted Foley catheters shown by scanning electron microscopyUrological Research, 1989
- Blockage of urinary catheters: Role of microorganisms and constituents of the urine on formation of encrustationsJournal of Clinical Epidemiology, 1989