Strategy for Safe Performance of Extrathoracic Magnetic Resonance Imaging at 1.5 Tesla in the Presence of Cardiac Pacemakers in Non–Pacemaker-Dependent Patients
Open Access
- 19 September 2006
- journal article
- clinical trial
- Published by Ovid Technologies (Wolters Kluwer Health) in Circulation
- Vol. 114 (12), 1285-1292
- https://doi.org/10.1161/circulationaha.105.597013
Abstract
Background— The purpose of the present study was to evaluate a strategy for safe performance of extrathoracic magnetic resonance imaging (MRI) in non–pacemaker-dependent patients with cardiac pacemakers. Methods and Results— Inclusion criteria were presence of a cardiac pacemaker and urgent clinical need for an MRI examination. Pacemaker-dependent patients and those requiring examinations of the thoracic region were excluded. The study group consisted of 82 pacemaker patients who underwent a total of 115 MRI examinations at 1.5T. To minimize radiofrequency-related lead heating, the specific absorption rate was limited to 1.5 W/kg. All pacemakers were reprogrammed before MRI: If heart rate was 60 bpm, sense-only mode was used to avoid MR-induced competitive pacing and potential proarrhythmia. Patients were monitored with ECG and pulse oximetry. All pacemakers were interrogated immediately before and after the MRI examination and after 3 months, including measurement of pacing capture threshold (PCT) and serum troponin I levels. All MR examinations were completed safely. Inhibition of pacemaker output or induction of arrhythmias was not observed. PCT increased significantly from pre- to post-MRI ( P =0.017). In 2 of 195 leads, an increase in PCT was only detected at follow-up. In 4 of 114 examinations, troponin increased from a normal baseline value to above normal after MRI, and in 1 case (troponin pre-MRI 0.02 ng/mL, post-MRI 0.16 ng/mL), this increase was associated with a significant increase in PCT. Conclusions— Extrathoracic MRI of non–pacemaker-dependent patients can be performed with an acceptable risk-benefit ratio under controlled conditions and by taking both MR- and pacemaker-related precautions.Keywords
This publication has 21 references indexed in Scilit:
- Government Viewpoint: U.S. Food & Drug Administration: Pacemakers, ICDs and MRIPacing and Clinical Electrophysiology, 2005
- MRT-Untersuchungen des Hirns bei Patienten mit implantierten Herzschrittmachern. Experimentelle und klinische Untersuchungen an einem 1,5 Tesla-SystemRöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 2005
- In vivo heating of pacemaker leads during magnetic resonance imagingEuropean Heart Journal, 2004
- Markers of Myocardial Damage, Tissue Healing, and Inflammation After Radiofrequency Catheter Ablation of Atrial TachyarrhythmiasJournal of Cardiovascular Electrophysiology, 2004
- Pacemaker Reed Switch Behavior in 0.5, 1.5, and 3.0 Tesla Magnetic Resonance Imaging Units: Are Reed Switches Always Closed in Strong Magnetic Fields?Pacing and Clinical Electrophysiology, 2002
- Interference with Cardiac Pacemakers by Magnetic Resonance Imaging: Are There Irreversible Changes at 0.5 Tesla?Pacing and Clinical Electrophysiology, 2001
- Effects of magnetic resonance imaging on cardiac pacemakers and electrodesAmerican Heart Journal, 1997
- Transesophageal cardiac pacing during magnetic resonance imaging: Feasibility and safety considerationsMagnetic Resonance in Medicine, 1996
- Estimating radiofrequency power deposition in body NMR imagingMagnetic Resonance in Medicine, 1985
- Changes in myocardial threshold. Physiologic and pharmacologic factors in patients with implanted pacemakersAmerican Heart Journal, 1967