Intelligent stretching of ankle joints with contracture/spasticity
- 1 September 2002
- journal article
- clinical trial
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Neural Systems and Rehabilitation Engineering
- Vol. 10 (3), 149-157
- https://doi.org/10.1109/tnsre.2002.802857
Abstract
An intelligent stretching device was developed to treat the spastic/contractured ankle of neurologically impaired patients. The device stretched the ankle safely throughout the range of motion (ROM) to extreme dorsiflexion and plantarflexion until a specified peak resistance torque was reached with the stretching velocity controlled based on the resistance torque. The ankle was held at the extreme position for a period of time to let stress relaxation occur before it was rotated back to the other extreme position. Stretching was slow at the joint extreme positions, making it possible to reach a larger ROM safely and it was fast in the middle ROM so the majority of the treatment was spent in stretching the problematic extreme ROM. Furthermore, the device evaluated treatment outcome quantitatively in multiple aspects, including active and passive ROM, joint stiffness and viscous damping and reflex excitability. The stretching resulted in considerable changes in joint passive ROM, stiffness, viscous damping and reflex gain. The intelligent control and yet simple design of the device suggest that with appropriate simplification, the device can be made portable and low cost, making it available to patients and therapists for frequent use in clinics/home and allowing more effective treatment and long-term improvement.Keywords
This publication has 22 references indexed in Scilit:
- Dynamic and static control of the human knee joint in abduction–adductionJournal of Biomechanics, 2001
- Stiffness, viscosity, and upper‐limb inertia about the glenohumeral abduction axisJournal of Orthopaedic Research, 2000
- Reflex hyperexcitability and muscle contracture in relation to spastic hypertoniaCurrent Opinion in Neurology, 1996
- Spasticity and muscle contracture following strokeBrain, 1996
- Technique forin vivo measurement of the three-dimensional kinematics and laxity characteristics of the ankle joint complexJournal of Orthopaedic Research, 1994
- The Biologic Concept of Continuous Passive Motion of Synovial JointsClinical Orthopaedics and Related Research, 1989
- FOR HOW LONG MUST THE SOLEUS MUSCLE BE STRETCHED EACH DAY TO PREVENT CONTRACTURE?Developmental Medicine and Child Neurology, 1988
- Effect of muscular activity on valgus/varus laxity and stiffness of the kneeJournal of Biomechanics, 1986
- Rehabilitation for stroke: a review.Stroke, 1986
- Experimental rapid sarcomere loss with concomitant hypoextensibilityMuscle & Nerve, 1981