Corneal Thickness- and Age-Related Biomechanical Properties of the Cornea Measured with the Ocular Response Analyzer

Abstract
Purpose. The Ocular Response Analyzer (ORA) is a new instrument that measures the corneal biomechanical response (corneal hysteresis, CH) to rapid indentation by an air jet. CH is the difference in applanation pressures (P1, P2) between the rising and falling phases of the air jet. The investigation had two parts: a characterization study and a validation study. In the characterization study, the purposes were to investigate the intraocular pressure (IOP)–dependence of CH and to characterize the performance of the ORA. In the validation study, the purposes were to investigate the association between CH and both age and central corneal thickness (CCT) and the agreement between ORA and Goldmann applanation tonometer (GAT) IOP measurements.