This report describes the current state of flattening filter-free (FFF) radiotherapy beams implemented on conventional linear accelerators, and is aimed primarily at practicing medical physicists. The Therapy Emerging Technology Assessment Work Group of the American Association of Physicists in Medicine (AAPM) formed a writing group to assess FFF technology. The published literature on FFF technology was reviewed, along with technical specifications provided by vendors. Based on this information, supplemented by the clinical experience of the group members, consensus guidelines and recommendations for implementation of FFF technology were developed. Areas in need of further investigation were identified. Removing the flattening filter increases beam intensity, especially near the central axis. Increased intensity reduces treatment time, especially for high-dose stereotactic radiotherapy/radiosurgery (SRT/SRS). Furthermore, removing the flattening filter reduces out-of-field dose and improves beam modeling accuracy. FFF beams are advantageous for small field (e.g., SRS) treatments and are appropriate for intensity-modulated radiotherapy (IMRT). For conventional 3D radiotherapy of large targets, FFF beams may be disadvantageous compared to flattened beams because of the heterogeneity of FFF beam across the target (unless modulation is employed). For any application, the nonflat beam characteristics and substantially higher dose rates require consideration during the commissioning and quality assurance processes relative to flattened beams, and the appropriate clinical use of the technology needs to be identified. Consideration also needs to be given to these unique characteristics when undertaking facility planning. Several areas still warrant further research and development. Recommendations pertinent to FFF technology, including acceptance testing, commissioning, quality assurance, radiation safety, and facility planning, are presented. Examples of clinical applications are provided. Several of the areas in which future research and development are needed are also indicated. PACS number: 87.53.-j, 87.53.Bn, 87.53.Ly, 87.55.-x, 87.55.N-, 87.56.bcKeywords
This publication has 98 references indexed in Scilit:
- Credentialing results from IMRT irradiations of an anthropomorphic head and neck phantomMedical Physics, 2013
- Volumetric Modulation Arc Radiotherapy With Flattening Filter-Free Beams Compared With Static Gantry IMRT and 3D Conformal Radiotherapy for Advanced Esophageal Cancer: A Feasibility StudyInternational Journal of Radiation Oncology*Biology*Physics, 2012
- The Use of Photon Beams of a Flattening Filter-free Linear Accelerator for Hypofractionated Volumetric Modulated Arc Therapy in Localized Prostate CancerInternational Journal of Radiation Oncology*Biology*Physics, 2012
- Surface Dose Investigation of the Flattening Filter-Free Photon BeamsInternational Journal of Radiation Oncology*Biology*Physics, 2012
- Feasibility and early clinical assessment of flattening filter free (FFF) based stereotactic body radiotherapy (SBRT) treatmentsRadiation Oncology, 2011
- Lowering Whole-Body Radiation Doses in Pediatric Intensity-Modulated Radiotherapy Through the Use of Unflattened Photon BeamsInternational Journal of Radiation Oncology*Biology*Physics, 2011
- Dose-Escalated Radiotherapy for High-Risk Prostate Cancer: Outcomes in Modern Era With Short-Term Androgen Deprivation TherapyInternational Journal of Radiation Oncology*Biology*Physics, 2010
- Evaluation of Safety in a Radiation Oncology Setting Using Failure Mode and Effects AnalysisInternational Journal of Radiation Oncology*Biology*Physics, 2009