Intraocular metallic-appearing foreign bodies after phacoemulsification
- 1 November 1996
- journal article
- Published by Ovid Technologies (Wolters Kluwer Health) in Journal of Cataract & Refractive Surgery
- Vol. 22 (9), 1247-1250
- https://doi.org/10.1016/s0886-3350(96)80078-2
Abstract
Intraocular metallic-appearing foreign bodies in the anterior segment were seen in six eyes after cataract extraction by two-handed phacoemulsification. These fragments produced no symptoms and had no clinically significant effect on visual function. Similar-appearing foreign bodies were experimentally produced by applying ultrasound power with a phacoemulsification handpiece to the surface of a cyclodialysis spatula in vitro. Microscopic examination of the phaco handpiece tip and cyclodialysis spatulas revealed numerous surface irregularities, suggesting a possible source of these fragments. These results suggest that instrument touch during phacoemulsification may be the cause of the fragments. Long-term follow-up of these patients is required to determine the overall effect of the fragments.Keywords