Left Ventricular Assist Device Outflow Graft Obstruction: A Complication Specific to Polytetrafluoroethylene Covering. A Word of Caution!
- 1 August 2019
- journal article
- research article
- Published by Ovid Technologies (Wolters Kluwer Health) in ASAIO Journal
- Vol. 65 (6), e58-e62
- https://doi.org/10.1097/mat.0000000000000929
Abstract
Pump thrombosis is an established complication of left ventricular assist devices (LVADs). Outflow graft obstruction has been previously described as one cause of LVAD thrombosis. We identified four cases of outflow graft obstruction that were attributed to a commonly applied polytetrafluoroethylene (PTFE) covering of the outflow graft. In this set of patients, the outflow graft was obstructed by a thrombus which formed between the outflow graft and its external PTFE covering, leading to impingement of the outflow graft. Patients typically presented after a median duration of 26 months (range 23–41 months) of support with gradual increase of heart failure symptoms and low pump flows. Computed tomography angiography was found to be the best diagnostic modality. Treatments included surgical LVAD replacement as well as percutaneous intraluminal stenting of the outflow graft. Our findings indicate that PTFE graft covering of the LVAD outflow graft can lead to graft occlusion and should be reconsidered as a potentially harmful modification to the approved device implant technique.Keywords
This publication has 21 references indexed in Scilit:
- Left Ventricular Assist Device Inflow Cannula Angle and Thrombosis RiskCirculation: Heart Failure, 2018
- Platelet activation is a preoperative risk factor for the development of thromboembolic complications in patients with continuous‐flow left ventricular assist deviceEuropean Journal of Heart Failure, 2017
- Intermittent Aortic Valve Opening and Risk of Thrombosis in Ventricular Assist Device PatientsASAIO Journal, 2017
- LVAD Outflow Graft Angle and Thrombosis RiskASAIO Journal, 2017
- PREVENtion of HeartMate II Pump Thrombosis Through Clinical Management: The PREVENT multi-center studyThe Journal of Heart and Lung Transplantation, 2016
- Diagnosis and Treatment Algorithm for Blood Flow Obstructions in Patients With Left Ventricular Assist DeviceJournal of the American College of Cardiology, 2016
- Science for surgeons: Understanding pump thrombogenesis in continuous-flow left ventricular assist devicesThe Journal of Thoracic and Cardiovascular Surgery, 2015
- Advanced heart failure in patients infected with human immunodeficiency virus: Is there equal access to care?The Journal of Heart and Lung Transplantation, 2014
- An analysis of pump thrombus events in patients in the HeartWare ADVANCE bridge to transplant and continued access protocol trialThe Journal of Heart and Lung Transplantation, 2013
- Surgical Technique Influences HeartMate II Left Ventricular Assist Device ThrombosisThe Annals of Thoracic Surgery, 2013