Monitoring of end tidal carbon dioxide and transcutaneous carbon dioxide during neonatal transport
Open Access
- 7 June 2005
- journal article
- research article
- Published by BMJ in Archives of Disease in Childhood: Fetal & Neonatal
- Vol. 90 (6), F523-F526
- https://doi.org/10.1136/adc.2004.064717
Abstract
Objective: To assess the accuracy of measurements of end tidal carbon dioxide (CO2) during neonatal transport compared with arterial and transcutaneous measurements. Design: Paired end tidal and transcutaneous CO2 recordings were taken frequently during road transport of 21 ventilated neonates. The first paired CO2 values were compared with an arterial blood gas. The differences between arterial CO2 (Paco2), transcutaneous CO2 (TcPco2), and end tidal CO2 (Petco2) were analysed. The Bland-Altman method was used to assess bias and repeatability. Results: Petco2 correlated strongly with Paco2 and TcPco2. However, Petco2 underestimated Paco2 at a clinically unacceptable level (mean (SD) 1.1 (0.70) kPa) and did not trend reliably over time within individual subjects. The Petco2 bias was independent of Paco2 and severity of lung disease. Conclusions: Petco2 had an unacceptable under-recording bias. TcPco2 should currently be considered the preferred method of non-invasive CO2 monitoring for neonatal transport.Keywords
This publication has 19 references indexed in Scilit:
- Transcutaneous measurement of carbon dioxide tension during long-distance transport of neonates receiving mechanical ventilation.1998
- Mainstream End-tidal Carbon Dioxide Monitoring in the Neonatal Intensive Care UnitPEDIATRICS, 1998
- Evaluation of a Portable Infrared End-Tidal Carbon Dioxide Monitor During Pediatric Interhospital TransportPEDIATRICS, 1995
- Estimation of arterial carbon dioxide by end-tidal and transcutaneous PCO2 measurements in ventilated childrenPediatric Pulmonology, 1992
- End-tidal carbon dioxide pressure in neonates and infants measured by aspiration and flow-through capnographyJournal of Clinical Monitoring and Computing, 1991
- Continuous end-tidal CO2 sampling within the proximal endotracheal tube estimates arterial CO2 tension in infantsCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 1991
- Is Distal Sampling of End-tidal CO2Necessary in Small Subjects?Anesthesiology, 1990
- Accuracy of end-tidal PCO2 measurements using a sidestream capnometer in infants and children ventilated with the Sechrist infant ventilatorCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 1990
- Transcutaneous carbon dioxide monitoring during neonatal transportCritical Care Medicine, 1986
- Statistical methods for assessing agreement between two methods of clinical measurement.1986