An Outbreak of Gram-Negative Bacteremia in Hemodialysis Patients Traced to Hemodialysis Machine Waste Drain Ports

Abstract
Objective: To investigate an outbreak of gram-negative bacteremias at a hemodialysis center (December 1, 1996-January 31, 1997).Design: Retrospective cohort study. Reviewed infection control practices and maintenance and disinfection procedures for the water system and dialysis machines. Performed cultures of the water and dialysis machines, including the waste-handling option (WHO), a drain port designed to dispose of saline used to flush the dialyzer before patient use. Compared isolates by pulsed-field gel electrophoresis.Setting: A hemodialysis center in Maryland.Results: 94 patients received dialysis on 27 machines; 10 (11%) of the patients had gram-negative bacteremias. Pathogens causing these infections wereEnterobacter cloacae(n=6),Pseudomonas aeruginosa(n=4), andEscherichia coli(n=2); two patients had polymicrobial bacteremia. Factors associated with development of gram-negative bacteremias were receiving dialysis via a central venous catheter (CVC) rather than via an arterio-venous shunt (all 10 infected patients had CVCs compared to 31 of 84 uninfected patients, relative risk [RR] undefined;P<.001) or dialysis on any of three particular dialysis machines (7 of 10 infected patients were exposed to the three machines compared to 20 of 84 uninfected patients, RR=5.8;P=.005).E cloacae, P aeruginosa, or both organisms were grown from cultures obtained from several dialysis machines. WHO valves, which prevent backflow from the drain to dialysis bloodlines, were faulty in 8 (31%) of 26 machines, including 2 of 3 machines epidemiologically linked to case-patients. Pulsed-field gel electrophoresis patterns of available dialysis machine and patientE cloacaeisolates were identical.Conclusions: Our study suggests that WHO ports with incompetent valves and resultant backflow were a source of cross-contamination of dialysis bloodlines and patients' CVCs. Replacement of faulty WHO valves and enhanced disinfection of dialysis machines terminated the outbreak.