Electrocardiographic Motion Artifact Versus Electrode Impedance
- 20 December 2004
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. 52 (1), 136-139
- https://doi.org/10.1109/tbme.2004.836503
Abstract
We degraded electrocardiographic electrodes by exposing them to air for four days and evaluated them on 12 subjects. After application, we recorded the electrocardiogram (including motion artifact), missed QRS detections and electrode impedance during 5 min of arm and body movements. Missed QRS detections increased with electrode impedance but correlation was poor. Increased electrode impedance was not a reliable predictor of a poor electrode and the need to replace it.Keywords