Comparison of the GlideScope video laryngoscope and Macintosh laryngoscope in simulated tracheal intubation scenarios
- 1 May 2008
- journal article
- research article
- Published by BMJ in Emergency Medicine Journal
- Vol. 25 (5), 279-282
- https://doi.org/10.1136/emj.2007.052803
Abstract
Objective: To compare the GlideScope video laryngoscope (GVL) with the classic Macintosh laryngoscope in simulated airway scenarios of varying difficulty. Materials and methods: A prospective, crossover and randomised study was performed. Four airway scenarios were simulated using the Airsim model as follows: normal; cervical spine immobilisation; tongue oedema and combined cervical spine immobilisation with tongue oedema. Emergency physicians performed tracheal intubations using both devices in each of the scenarios. The time required to intubate, the success rate and the number of intubation attempts were recorded. At the end of each scenario, participants scored vocal cord visualisation using the percentage of glottic opening (POGO) visible and the subjective ease of intubation on a visual analogue scale (VAS). Results: All 25 participants successfully completed the study. There was no difference in the time required for successful tracheal intubation using the GVL compared with using the Macintosh laryngoscope in the four airway scenarios. Only one participant failed to intubate the trachea with the Macintosh laryngoscope for the combined scenario. There was a significant increase in POGO when using the GVL in the cervical spine immobilisation group (p = 0.027). The VAS score of the subjective ease of intubation was lower for the GVL than for the Macintosh laryngoscope device in difficult scenarios but this difference was not significant. Conclusion: This study suggests that the GVL could be an option for airway management even by emergency physicians with little experience and no training in its use.Keywords
This publication has 18 references indexed in Scilit:
- Comparison of haemodynamic responses to orotracheal intubation with GlideScope® videolaryngoscope and fibreoptic bronchoscopeEuropean Journal of Anaesthesiology, 2006
- A randomised, crossover study of the Dorges, McCoy and Macintosh laryngoscope blades in a simulated difficult intubation scenarioAnaesthesia, 2006
- Evaluation of intubation using the Airtraq® or Macintosh laryngoscope by anaesthetists in easy and simulated difficult laryngoscopy – a manikin study*Anaesthesia, 2006
- Potential cervical spine injury and difficult airway management for emergency intubation of trauma adults in the emergency department--a systematic reviewEmergency Medicine Journal, 2006
- Airway Management by US and Canadian Emergency Medicine Residents: A Multicenter Analysis of More Than 6,000 Endotracheal Intubation AttemptsAnnals of Emergency Medicine, 2005
- Practice Guidelines for Management of the Difficult AirwayAnesthesiology, 2003
- The failed intubation attempt in the Emergency Department: analysis of prevalence, rescue techniques, and personnelThe Journal of Emergency Medicine, 2002
- A grading system for direct laryngoscopyAnaesthesia, 1999
- Rapid‐sequence Intubation at an Emergency Medicine Residency: Success Rate and Adverse Events during a Two‐year PeriodAcademic Emergency Medicine, 1999
- Airway Management in the Emergency Department: A One-Year Study of 610 Tracheal IntubationsAnnals of Emergency Medicine, 1998