Background Current visual field screening machines are bulky and expensive, limiting their accessibility, affordability and use. We report the design and evaluation of a novel, portable, cost-effective system for glaucoma screening using smartphone-based visual field screening using frequency doubling technology (FDT) and a head-mounted display. Methods Nineteen eyes of 10 subjects with new-onset or chronic primary open angle glaucoma were tested and compared with the Humphrey Zeiss FDT and the newly developed Mobile Virtual Perimetry (MVP) FDT with the C-20 testing pattern. Mann-Whitney, Bland-Altman and linear regression analyses were performed to assess statistical difference, agreement and correlation, respectively, between the two devices. Results The average age of the participants was 58±15 years. No statistically significant difference was found between the MVP FDT and the Humphrey Zeiss FDT (p>0.05). Bland-Altman and linear regression analyses demonstrated good agreement and correlation between the two devices. Conclusion The MVP FDT is a low-cost, portable visual field screening device that produces comparable results to the Humphrey Zeiss FDT and may be used as an easily accessible screening tool for glaucoma.Keywords
Funding Information
- Walter G. Ross Foundation
- Research to Prevent Blindness
- NIH Center Core Grant (P30EY014801)
This publication has 16 references indexed in Scilit:
- Glaucoma Screening in the Haitian Afro-Caribbean Population of South FloridaPLOS ONE, 2014
- MultiRankSeq: Multiperspective Approach for RNAseq Differential Expression Analysis and Quality ControlBioMed Research International, 2014
- Testing of Visual Field with Virtual Reality Goggles in Manual and Visual Grasp ModesBioMed Research International, 2014
- Characterization of the Opp Peptide Transporter ofCorynebacterium pseudotuberculosisand Its Role in Virulence and PathogenicityBioMed Research International, 2014
- Screening for Glaucoma: U.S. Preventive Services Task Force Recommendation StatementAnnals of Internal Medicine, 2013
- Screening for glaucomaCurrent Opinion in Opthalmology, 2013
- Detectability of Glaucomatous Changes Using SAP, FDT, Flicker Perimetry, and OCTJournal of Glaucoma, 2009
- The Influence of Learning Effect on Frequency Doubling Technology Perimetry (Matrix)Journal of Glaucoma, 2007
- The number of people with glaucoma worldwide in 2010 and 2020British Journal of Ophthalmology, 2006
- Neuronal death in glaucomaProgress in Retinal and Eye Research, 1999