Artifacts in CT: Recognition and Avoidance
Top Cited Papers
- 1 November 2004
- journal article
- Published by Radiological Society of North America (RSNA) in RadioGraphics
- Vol. 24 (6), 1679-1691
- https://doi.org/10.1148/rg.246045065
Abstract
Artifacts can seriously degrade the quality of computed tomographic (CT) images, sometimes to the point of making them diagnostically unusable. To optimize image quality, it is necessary to understand why artifacts occur and how they can be prevented or suppressed. CT artifacts originate from a range of sources. Physics-based artifacts result from the physical processes involved in the acquisition of CT data. Patient-based artifacts are caused by such factors as patient movement or the presence of metallic materials in or on the patient. Scanner-based artifacts result from imperfections in scanner function. Helical and multisection technique artifacts are produced by the image reconstruction process. Design features incorporated into modern CT scanners minimize some types of artifacts, and some can be partially corrected by the scanner software. However, in many instances, careful patient positioning and optimum selection of scanning parameters are the most important factors in avoiding CT artifacts.Keywords
This publication has 3 references indexed in Scilit:
- Generalized multi‐dimensional adaptive filtering for conventional and spiral single‐slice, multi‐slice, and cone‐beam CTMedical Physics, 2001
- Artefacts in spiral-CT images and their relation to pitch and subject morphologyEuropean Radiology, 1999
- Algorithm for image reconstruction in multi-slice helical CTMedical Physics, 1998