Isotope-Ratio-Monitoring Liquid Chromatography Mass Spectrometry (IRM-LCMS): First Results from a Moving Wire Interface System
- 1 August 1996
- journal article
- research article
- Published by Taylor & Francis Ltd in Isotopes in Environmental and Health Studies
- Vol. 32 (2-3), 275-283
- https://doi.org/10.1080/10256019608036320
Abstract
A Liquid Chromatography-Combustion (LC-C) Interface, based on a moving wire technique, has been built and tested. The LC effluent is deposited onto a transport wire, which carries the sample through solvent evaporation and combustion ovens. CO2 from the combustion step is analysed in an isotope ratio mass spectrometer. Performance of the interface was tested by loop injections of sucrose and glucose into a liquid flow of methanol/water (80/20). Accuracy and precision of δ13CPDB < 1‰ was achieved for sample concentrations > 500 ng/ul (5μl loop), sufficient for studies at natural isotope ratios. In case of 13C tracer applications the detection limit was determined to be about 20 pg carbon tracer (on wire).Keywords