Evaluation of NucliSens easyMAG for Automated Nucleic Acid Extraction from Various Clinical Specimens

Abstract
The objectives of this study were to evaluate the performance of the NucliSens easyMAG platform for nucleic acid extraction from different clinical specimens compared to NucliSens miniMAG platform and manual QIAGEN extraction. The NucliSens easyMAG and the NucliSens miniMAG showed equal performance on 215 throat swabs since real-time nucleic acid sequence-based amplification scored the same samples positive for Mycoplasma pneumoniae ( n = 9) and Chlamydia pneumoniae ( n = 5) RNAs, although internal control RNA was slightly better detected with the NucliSens easyMAG (99.3% versus 96.8%). NucliSens easyMAG extracted nucleic acids more efficiently (higher recovery and/or fewer inhibitors) compared to QIAGEN extraction by showing, on average, lower Ct values in real-time LightCycler PCR, although 4 individual specimen out of 45 were found positive only with QIAGEN. For nine M. pneumoniae -positive throat swabs, the mean difference in Ct values between NucliSens easyMAG extraction and QIAGEN extraction was −2.26 (range, −5.77 to +0.60); for the detection of five C. pneumoniae -positive throat swabs, the average difference in Ct values between the two methods was −3.38 (range, −6.62 to −2.02); and for the detection of cytomegalovirus in 24 blood samples, the mean difference in Ct values between the two methods was −0.95 (range, −5.51 to +1.68). The NucliSens easyMAG is considerably easier to perform, efficiently extracts nucleic acids from throat swabs and whole blood, is automated, and has high throughput.