Feasibility, safety, and efficacy of real-time three-dimensional transoesophageal echocardiography for guiding device closure of interatrial communications: initial clinical experience and impact on radiation exposure

Abstract
Our aim was to assess the feasibility and safety of real-time (RT) three-dimensional (3D) transoesophageal echocardiography (TEE) for guiding transcatheter closure of interatrial communications and to evaluate its additional benefit over conventional 2D TEE in reducing radiation exposure for the patient. Twenty-five patients undergoing device closure of their interatrial defect had the procedure guided by fluoroscopy, 2D TEE, and RT 3D TEE. We retrospectively compared this group with a historical control group in which interventional guidance was performed using fluoroscopy and 2D TEE alone. The application of RT 3D TEE allowed safe device deployment in all patients without any complications, resulting in a reduction of mean fluoroscopy time (10 ± 6 to 6 ± 4 min, P < 0.01), mean dose area product (DAP) (964 ± 628 to 535 ± 464 cGy cm 2 , P < 0.01), and mean DAP per individual body surface area (494 ± 317 to 273 ± 221 cGy cm 2 /m 2 , P < 0.01). RT 3D TEE as an adjunct to 2D TEE is a feasible and safe tool to guide transcatheter device closure of interatrial communications, resulting in a reduction of radiation exposure. These data indicate that RT 3D TEE can be used to safely monitor interatrial defect closure in clinical routine.

This publication has 18 references indexed in Scilit: