Feasibility of using finger arterial pressure in neonates
Open Access
- 1 September 1997
- journal article
- research article
- Published by BMJ in Archives of Disease in Childhood: Fetal & Neonatal
- Vol. 77 (2), F139-F140
- https://doi.org/10.1136/fn.77.2.f139
Abstract
The feasibility of using a Finapres device to reproduce the beat to beat signal of arterial blood pressure in eight neonates was assessed and compared with intra-arterial measurement of arterial blood pressure in the umbilical artery, using a catheter. The two methods gave similar results. Continuous recording of arterial blood pressure in neonates using Finapres is feasible and reliable.Keywords