An assessment of the Dinamap 845
Open Access
- 1 March 1984
- journal article
- research article
- Published by Wiley in Anaesthesia
- Vol. 39 (3), 261-267
- https://doi.org/10.1111/j.1365-2044.1984.tb07240.x
Abstract
The accuracy of the Dinamap 845 automatic blood pressure recorder was assessed by comparing its own indirect determinations of blood pressure with direct intra-arterial recordings. It was found that in the majority of cases it was capable of producing reliable trend information during anaesthesia. The instrument may not be able to interpret pressure signals from a patient with a severe dysrhythmia. It is probably an unsuitable monitor for use with very rapidly acting drugs such as sodium nitroprusside.Keywords
This publication has 11 references indexed in Scilit:
- Noninvasive estimation of central aortic pressure using the oscillometric method for analyzing systemic artery pulsatile blood flow: Comparative study of indirect systolic, diastolic, and mean brachial artery pressure with simultaneous direct ascending aortic pressure measurementsAmerican Heart Journal, 1982
- Indirect Measurement of Blood Pressure in Neonates and Infants Utilizing an Automatic Noninvasive Oscillometric MonitorAnesthesia & Analgesia, 1981
- Ulnar Nerve Palsy Possibly Related to Use of Automatically Cycled Blood Pressure CuffAnesthesia & Analgesia, 1981
- An Automated Oscillometric Technique for Estimating Mean Arterial Pressure in Critically III NewbornsAnesthesiology, 1981
- Clinical evaluation of Dinamap 845 automated blood pressure recorder.Heart, 1980
- The Meaning of the Point of Maximum Oscillations in Cuff Pressure in the Indirect Measurement of Blood Pressure—Part IIJournal of Biomechanical Engineering, 1980
- Indirect Measurement of Mean Blood Pressure in the Anesthetized PatientAnesthesiology, 1979
- Comparison of Indirect and Direct Methods of Measuring Arterial Blood PressureCell Metabolism, 1954