Temperature dependence of commercially available diode detectors

Abstract
Temperature dependence of commercially available n- and p-type diodes were studied experimentally under both high instantaneous dose rate (pulsed) and low dose rate (continuous) radiation. The sensitivity versus temperature was measured at SSD = 80 or 100 cm, 10 x 10 cm2, and 5 cm depth in a 30 x 30 x 30 cm3 water phantom between 10 degrees C and 35 degrees C. The response was linear for all the diode detectors. The temperature coefficient (or sensitivity variation with temperature, svwt) was dose rate independent for preirradiated diodes. They were (0.30 +/- 0.01)%/degrees C, (0.36 +/- 0.03)%/degrees C, and (0.29 +/- 0.08)%/degrees C for QED p-type, EDP p-type, and Isorad n-type diodes, respectively. The temperature coefficient for unirradiated n-type diodes was different under low dose rate [(0.16 to 0.45)%/degrees C, continuous, cobalt] and high instantaneous dose rate [(0.07 +/- 0.02)%/degrees C, pulsed radiation]. Moreover, the temperature coefficient varies among individual diodes. Similarly, the temperature coefficient for a special unirradiated QED p-type diode was different under low dose rate (0.34%/degrees C, cobalt) and high instantaneous dose rate [(0.26 +/- 0.01)%/degrees C, pulsed radiation]. Sufficient preirradiation can eliminate dose rate dependence of the temperature coefficient. On the contrary, preirradiation cannot eliminate dose rate dependence of the diode sensitivity itself.