Realistic simulation of reduced-dose CT with noise modeling and sinogram synthesis using DICOM CT images
- 4 December 2013
- journal article
- Published by Wiley in Medical Physics
- Vol. 41 (1), 011901
- https://doi.org/10.1118/1.4830431
Abstract
Reducing the patient dose while maintaining the diagnostic image quality during CT exams is the subject of a growing number of studies, in which simulations of reduced-dose CT with patient data have been used as an effective technique when exploring the potential of various dose reduction techniques. Difficulties in accessing raw sinogram data, however, have restricted the use of this technique to a limited number of institutions. Here, we present a novel reduced-dose CT simulation technique which provides realistic low-dose images without the requirement of raw sinogram data. Two key characteristics of CT systems, the noise equivalent quanta (NEQ) and the algorithmic modulation transfer function (MTF), were measured for various combinations of object attenuation and tube currents by analyzing the noise power spectrum (NPS) of CT images obtained with a set of phantoms. Those measurements were used to develop a comprehensive CT noise model covering the reduced x-ray photon flux, object attenuation, system noise, and bow-tie filter, which was then employed to generate a simulated noise sinogram for the reduced-dose condition with the use of a synthetic sinogram generated from a reference CT image. The simulated noise sinogram was filtered with the algorithmic MTF and back-projected to create a noise CT image, which was then added to the reference CT image, finally providing a simulated reduced-dose CT image. The simulation performance was evaluated in terms of the degree of NPS similarity, the noise magnitude, the bow-tie filter effect, and the streak noise pattern at photon starvation sites with the set of phantom images. The simulation results showed good agreement with actual low-dose CT images in terms of their visual appearance and in a quantitative evaluation test. The magnitude and shape of the NPS curves of the simulated low-dose images agreed well with those of real low-dose images, showing discrepancies of less than +/-3.2% in terms of the noise power at the peak height and +∕-1.2% in terms of the spatial frequency at the peak height. The magnitudes of the noise measured for 12 different combinations the phantom size, tube current, and reconstruction kernel for the simulated and real low-dose images were very similar, with differences of 0.1 to 4.7%. The p value for a statistical testing of the difference in the noise magnitude ranged from 0.99 to 0.11, showing that there was no difference statistically between the noise magnitudes of the real and simulated low-dose images using our method. The strength and pattern of the streak noise in an anthropomorphic phantom was also consistent with expectations. A novel reduced-dose CT simulation technique was developed which uses only CT images while not requiring raw sinogram data. Our method can provide realistic simulation results under reduced-dose conditions both in terms of the noise magnitude and the textual appearance. This technique has the potential to promote clinical research for patient dose reductions.Keywords
Funding Information
- ICT and Future Planning, Korea (2013K000423)
- ICT and Future Planning, Korea (2013K000423)
This publication has 29 references indexed in Scilit:
- Projected Cancer Risks From Computed Tomographic Scans Performed in the United States in 2007Archives of Internal Medicine, 2009
- Computed Tomography — An Increasing Source of Radiation ExposureThe New England Journal of Medicine, 2007
- Properties of preprocessed sinogram data in x-ray computed tomographyMedical Physics, 2006
- Multi–Detector Row CT Pulmonary Angiography: Comparison of Standard-Dose and Simulated Low-Dose TechniquesRadiology, 2005
- An Extensible Software-based Platform for Reconstruction and Evaluation of CT ImagesRadioGraphics, 2004
- Computer-Simulated Radiation Dose Reduction for Abdominal Multidetector CT of Pediatric PatientsAmerican Journal of Roentgenology, 2002
- CT Colonography at Different Radiation Dose Levels: Feasibility of Dose ReductionRadiology, 2002
- Pulmonary Nodules: Experimental and Clinical Studies at Low-Dose CTRadiology, 1999
- Pulmonary nodule detection: low-dose versus conventional CT.Radiology, 1998
- Simulated dose reduction in conventional chest CT: validation study.Radiology, 1997