Validation of UV-VIS Spectrophotometric Methods for Determination of Inulin Levels from Lesser Yam (Dioscorea esculenta L.)
Open Access
- 9 August 2018
- journal article
- Published by Institute of Research and Community Services Diponegoro University (LPPM UNDIP) in Jurnal Kimia Sains dan Aplikasi
- Vol. 21 (4), 161-165
- https://doi.org/10.14710/jksa.21.4.161-165
Abstract
Validation of methods is necessary to ensure that the analytical methods used are appropriate and reliable. In this research, the determination of validation method and determination of inulin content from gembili bulb (Dioscorea esculenta L.). The method used for determining inulin levels is UV-Vis spectrophotometry. Based on the results obtained, the maximum wavelength determination of inulin levels is 520 nm. The method of analysis in this study meets the specified validation requirements. These parameters include the equation on the inulin raw curve y = 0.0045x +0.109 with the coefficient of determination 0.9992. The accuracy of the methods tested using the recovery test was 96.14-106.76% at concentrations of 25, 100, and 225 ppm. The intraday precision test is in the range of 2.67-6.82% at the same concentration as the accuracy test. Precision interday equal to 7.41% at concentration 225 ppm for 3 day analysis period. The values of LOD and LOQ are 3.58 ppm and 11.95 ppm, respectively. The concentration of inulin extract from gembili bulb is 840.44-12188.59 ppm.Keywords
This publication has 4 references indexed in Scilit:
- Inulin, a flexible oligosaccharide I: Review of its physicochemical characteristicsCarbohydrate Polymers, 2015
- KANDUNGAN INULIN DARI UMBI DAHLIASp YANG DITANAM PADA JENIS TANAH VERTISOL, INCEPTISOL DAN ANDISOLJurnal Kimia Terapan Indonesia, 2014
- POLYPHENOL AND INULIN CONTENT IN A COLLECTION OF ARTICHOKEActa Horticulturae, 2005
- PETUNJUK PELAKSANAAN VALIDASI METODE DAN CARA PERHITUNGANNYAMajalah Ilmu Kefarmasian, 2004