Time of Flight Mass Spectrometry Imaging of Samples Fractured In Situ with a Spring-Loaded Trap System
- 1 July 2010
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 82 (15), 6652-6659
- https://doi.org/10.1021/ac101243b
Abstract
An in situ freeze fracture device featuring a spring-loaded trap system has been designed and characterized for time of flight secondary ion mass spectrometry (TOF SIMS) analysis of single cells. The device employs the sandwich assembly, which is typically used in freeze fracture TOF SIMS experiments to prepare frozen, hydrated cells for high-resolution SIMS imaging. The addition of the spring-loaded trap system to the sandwich assembly offers two advances to this sample preparation method. First, mechanizing the fracture by adding a spring standardizes each fracture by removing the need to manually remove the top of the sandwich assembly with a cryogenically cooled knife. A second advance is brought about because the top of the sandwich is not discarded after the sandwich assembly has been fractured. This results in two imaging surfaces effectively doubling the sample size and providing the unique ability to image both sections of a cell bifurcated by the fracture. Here, we report TOF SIMS analysis of freeze fractured rat pheochromocytoma (PC12) cells using a Bi cluster ion source. This work exhibits the ability to obtain single cell chemical images with subcellular lateral resolution from cells preserved in an ice matrix. In addition to preserving the cells, the signal from lipid fragment ions rarely identified in single cells are better observed in the freeze-fractured samples for these experiments. Furthermore, using the accepted argument that K+ signal indicates a cell that has been fractured though the cytoplasm, we have also identified different fracture planes of cells over the surface. Coupling a mechanized freeze fracture device to high-resolution cluster SIMS imaging will provide the sensitivity and resolution as well as the number of trials required to carry out biologically relevant SIMS experiments.Keywords
This publication has 32 references indexed in Scilit:
- Mass spectrometry imaging of mating Tetrahymena show that changes in cell morphology regulate lipid domain formationProceedings of the National Academy of Sciences of the United States of America, 2010
- Fixation and Drying Protocols for the Preparation of Cell Samples for Time-of-Flight Secondary Ion Mass Spectrometry AnalysisAnalytical Chemistry, 2009
- A New Dynamic in Mass Spectral Imaging of Single Biological CellsAnalytical Chemistry, 2008
- Freeze-Etching and Vapor Matrix Deposition for ToF-SIMS Imaging of Single CellsLangmuir, 2008
- Spatially and Temporally Resolved Single-Cell Exocytosis Utilizing Individually Addressable Carbon Microelectrode ArraysAnalytical Chemistry, 2008
- The PC12 cell as model for neurosecretionActa Physiologica, 2007
- Phospholipid mediated plasticity in exocytosis observed in PC12 cellsBrain Research, 2007
- Secondary Ion MS Imaging To Relatively Quantify Cholesterol in the Membranes of Individual Cells from Differentially Treated PopulationsAnalytical Chemistry, 2007
- Molecule Specific Imaging of Freeze-Fractured, Frozen-Hydrated Model Membrane Systems Using Mass SpectrometryJournal of the American Chemical Society, 2000
- Die Analyse monomolekularer FestkörperoberflÄchenschichten mit Hilfe der SekundÄrionenemissionThe European Physical Journal A, 1970