Design and Calibration of Load Sensing Orthopaedic Implants
- 28 March 2008
- journal article
- Published by ASME International in Journal of Biomechanical Engineering
- Vol. 130 (2), 021009
- https://doi.org/10.1115/1.2898831
Abstract
Contact forces and moments act on orthopaedic implants such as joint replacements. The three forces and three moment components can be measured by six internal strain gauges and wireless telemetric data transmission. The accuracy of instrumented implants is restricted by their small size, varying modes of load transfer, and the accuracy of calibration. Aims of this study were to test with finite element studies design features to improve the accuracy, to develop simple but accurate calibration arrangements, and to select the best mathematical method for calculating the calibration constants. Several instrumented implants, and commercial and test transducers were calibrated using different loading setups and mathematical methods. It was found that the arrangement of flexible elements such as bellows or notches between the areas of load transfer and the central sensor locations is most effective to improve the accuracy. Increasing the rigidity of the implant areas, which are fixed in bones or articulate against joint surfaces, is less effective. Simple but accurate calibration of the six force and moment components can be achieved by applying eccentric forces instead of central forces and pure moments. Three different methods for calculating the measuring constants proved to be equally well suited. Employing these improvements makes it possible to keep the average measuring errors of many instrumented implants below 1–2% of the calibration ranges, including cross talk. Additional errors caused by noise of the transmitted signals can be reduced by filtering if this is permitted by the sampling rate and the required frequency content of the loads.Keywords
This publication has 26 references indexed in Scilit:
- Implantable 9-Channel Telemetry System for In Vivo Load Measurements With Orthopedic ImplantsIEEE Transactions on Biomedical Engineering, 2007
- An implantable telemetry device to measure intra-articular tibial forcesJournal of Biomechanics, 2005
- Treatment of open fractures of the shaft of the tibiaThe Journal of Bone and Joint Surgery, 2001
- Loads on internal spinal fixators measured in different body positionsEuropean Spine Journal, 1999
- Hip joint forces in sheepJournal of Biomechanics, 1999
- Patient monitoring system for load measurement with spinal fixation devicesMedical Engineering & Physics, 1996
- Torsional loads in the early postoperative period following total hip replacementJournal of Orthopaedic Research, 1995
- Influence of shoes and heel strike on the loading of the hip jointJournal of Biomechanics, 1995
- A six-degree-of-freedom transducer for in vitro measurement of patellofemoral contact forcesJournal of Biomechanics, 1994
- Telemeterized in vivo hip joint force data: A report on two patients after total hip surgeryJournal of Orthopaedic Research, 1991