Comparison of Flow Diversion and Coiling in Large Unruptured Intracranial Saccular Aneurysms

Abstract
Background and Purpose—: Flow diversion has emerged as an important tool for the management of intracranial aneurysms. The purpose of this study was to compare flow diversion and traditional embolization strategies in terms of safety, efficacy, and clinical outcomes in patients with unruptured, large saccular aneurysms (≥10 mm). Methods—: Forty patients treated with the Pipeline Embolization Device (PED) were matched in a 1:3 fashion with 120 patients treated with coiling based on patient age and aneurysm size. Fusiform and anterior communicating artery aneurysms were eliminated from the analysis. Procedural complications, angiographic results, and clinical outcomes were analyzed and compared. Results—: There were no differences between the 2 groups in terms of patient age, sex, aneurysm size, and aneurysm location. The rate of procedure-related complications did not differ between the PED (7.5%) and the coil group (7.5%; P =1). At the latest follow-up, a significantly higher proportion of aneurysms treated with PED (86%) achieved complete obliteration compared with coiled aneurysms (41%; P P P =0.8). Conclusions—: The PED provides higher aneurysm occlusion rates than coiling, with no additional morbidity and similar clinical outcomes. These findings suggest that the PED might be a preferred treatment option for large unruptured saccular aneurysms.