In Vitro Characterization of the eFlow Closed System Nebulizer with Glycopyrrolate Inhalation Solution
Open Access
- 1 June 2018
- journal article
- research article
- Published by Mary Ann Liebert Inc in Journal of Aerosol Medicine and Pulmonary Drug Delivery
- Vol. 31 (3), 162-169
- https://doi.org/10.1089/jamp.2017.1384
Abstract
Background: Glycopyrrolate administered by a novel, investigational eFlow® Closed System (CS) nebulizer (eFlow CS) is being evaluated for the maintenance treatment of chronic obstructive pulmonary disease (COPD). The eFlow CS is a hand-held, vibrating membrane nebulizer optimized to deliver 1 mL of glycopyrrolate solution into the lung in Methods: The aerosol performance of the eFlow CS nebulizer was characterized by delivered dose, aerodynamic droplet size distribution and nebulization time. Simulated use nebulizer performance over 60 days was assessed by volume median diameter (VMD), nebulized amount, and nebulization time. Nebulization outputs were assayed to ensure adequate delivery of glycopyrrolate with an acceptable impurity profile. Aerosol condensates were analyzed for glycopyrrolate concentration and impurities by ultra-high-performance liquid chromatography and compared with non-nebulized samples. Results: The mean mass median aerodynamic diameter, geometric standard deviation, and fine particle fraction were 3.7 μm, 1.7, and 72%, respectively, and independent of formulation strength (25 and 50 μg/mL). Delivered dose was 88% of the nominal dose for both formulation strengths. The mean delivered dose, assessed by breathing simulation, was 56.8% for 25 μg/mL and 62.6% for 50 μg/mL. Nebulization times were 1–2.5 minutes with no apparent increasing trend with use over a 60-day period. The nebulized amount showed no significant changes, whereas the VMD showed a slight, but not pharmaceutically relevant, increase (0.1–0.2 μm) after 60-day simulated use. Glycopyrrolate concentration and impurity levels of nebulized samples were statistically similar to those of non-nebulized samples. Conclusion: The eFlow CS generates glycopyrrolate aerosols with high delivered dose, short treatment time, and small droplet size with narrow size distribution suitable for central and peripheral airway deposition. The unit dose vial mitigates medication misuse and ensures dose uniformity. Results support the use of glycopyrrolate/eFlow CS for the treatment of COPD.Keywords
This publication has 11 references indexed in Scilit:
- Long-term safety of glycopyrrolate/eFlow® CS in moderate-to-very-severe COPD: Results from the Glycopyrrolate for Obstructive Lung Disease via Electronic Nebulizer (GOLDEN) 5 randomized studyRespiratory Medicine, 2017
- Efficacy and safety of glycopyrrolate/eFlow® CS (nebulized glycopyrrolate) in moderate-to-very-severe COPD: Results from the glycopyrrolate for obstructive lung disease via electronic nebulizer (GOLDEN) 3 and 4 randomized controlled trialsRespiratory Medicine, 2017
- Lung deposition of inhaled tobramycin with eFlow rapid/LC Plus jet nebuliser in healthy and cystic fibrosis subjectsJournal of Cystic Fibrosis, 2011
- Targeting Aerosolized Drugs to the Conducting Airways Using Very Large Particles and Extremely Slow InhalationsJournal of Aerosol Medicine and Pulmonary Drug Delivery, 2010
- The influence of breathing pattern during nebulization on the delivery of arformoterol using a breath simulator.2009
- Changes in Performance of the Pari eFlow® Rapid and Pari LC Plus™ during 6 Months Use by CF PatientsJournal of Aerosol Medicine and Pulmonary Drug Delivery, 2009
- Association between Norepinephrine Transporter Gene Polymorphism and Major DepressionNeuropsychobiology, 2004
- Performance comparison of nebulizer designs: constant-output, breath-enhanced, and dosimetric.2004
- Nebulizers: principles and performance.2000
- Comparison of twenty three nebulizer/compressor combinations for domiciliary useEuropean Respiratory Journal, 1995