An evaluation of the dosimetric performance characteristics of N-vinylpyrrolidone-based polymer gels
- 1 August 2007
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 52 (16), 5069-5083
- https://doi.org/10.1088/0031-9155/52/16/024
Abstract
The aim of this work was to investigate the dosimetric performance properties of theN-vinylpyrrolidone argon (VIPAR) based polymer gel as a dosimetric tool in clinical radiotherapy. VIPAR gels with a larger concentration of gelatin than the standard recipe were manufactured and irradiated up to 68 Gy using a 6 and 18 MV linear accelerator. Using MRI, the R2-dose response was recorded at different imaging sessions within a 34 day time period post-irradiation. The R2-dose response was found to be linear between 5 and 68 Gy. Although dose sensitivity did not show significant variation with time, the measured R2-dose values showed an increasing trend, which was less evident beyond 17 days. At one day post-irradiation, calculated dose standard uncertainties at 20 Gy and 56 Gy were 2.2% and 1.7%, providing a dose resolution of 0.45 Gy and 0.97 Gy, respectively. Although these values fulfilled the 2% limit of ICRU, when gels were imaged at one day post-irradiation, it was shown that the temporal evolution of the R2 values deteriorated the per cent standard uncertainty and the dose resolution by ∼57%, when imaged 17 days post-irradiation. Variation in the coagulation temperature of the gels did not impact the R2-dose sensitivity. This study has shown that the VIPAR gel has the properties of a dosimetric tool required in clinical radiotherapy, especially in applications where a wide dose dynamic range is employed. For results with the lowest per cent uncertainty and the optimum dose resolution, the dosimetry gels used in this work should be MR scanned at one day post-irradiation. Furthermore, a preliminary study on the R2-dose response of a new normoxicN-vinylpyrrolidone-based polymer gel showed that it could potentially replace the traditional VIPAR gel formulation, while preserving the wide dynamic dose response inherent to that monomer.This publication has 37 references indexed in Scilit:
- Myocardial and liver iron status using a fastT2* quantitative MRI (T2*qMRI) techniqueMagnetic Resonance in Medicine, 2007
- Polymer gel dosimeters with reduced toxicity: a preliminary investigation of the NMR and optical dose–response using different monomersPhysics in Medicine & Biology, 2006
- Investigation of tetrakis hydroxymethyl phosphonium chloride as an antioxidant for use in x-ray computed tomography polyacrylamide gel dosimetryPhysics in Medicine & Biology, 2006
- The fundamental radiation properties of normoxic polymer gel dosimeters: a comparison between a methacrylic acid based gel and acrylamide based gelsPhysics in Medicine & Biology, 2006
- Investigation of the PAGAT polymer gel dosimeter using magnetic resonance imagingPhysics in Medicine & Biology, 2005
- Dose verification of single shot gamma knife applications using VIPAR polymer gel and MRIPhysics in Medicine & Biology, 2005
- Essential characteristics of polymer gel dosimetersJournal of Physics: Conference Series, 2004
- Magnetic resonance imaging of radiation dose distributions using a polymer-gel dosimeterPhysics in Medicine & Biology, 1994
- Multiple‐spin‐echo imaging with a 2D Fourier methodMagnetic Resonance in Medicine, 1986
- Polymerization kinetics of polyacrylamide gels II. Effect of temperatureElectrophoresis, 1981