Prevention of Dislocation After Hip Arthroplasty

Abstract
Regarding dislocation after total hip arthroplasty, prevention is worth an ounce of cure. The current authors evaluated dislocation after total hip arthroplasty during the 26-year practice of one surgeon to identify potential variables that can contribute to the prevention of dislocation. Between 1970 and 1996, dislocation after total hip replacement occurred after 7.2% of primary hip arthroplasty procedures (298 of 4164 primary hip replacements) and 11.2% of revision hip arthroplasty procedures (90 of 803 revision hip replacements). Significant findings included an increase in dislocation when 22-mm modular femoral heads were used and a decrease in dislocation after revision for dislocation when constrained liners were used. An additional finding was that 26% of first time dislocations occurred more than 2 years after surgery. Concerning prevention of dislocation, small head modular femoral components should be used cautiously, and constrained liners should be considered in complex revision cases. Patients should be counseled concerning the potential for dislocation many years after their arthroplasty.