On the Accuracy of a New Displacement Instrumentation for Rotary Tablet Presses

Abstract
The Portable Press Analyzer™ (PPA; Puuman Oy, Finland), a commercially available instrumentation for rotary tablet presses, was tested for accuracy of determination of force and displacement. The calibration of the force transducers (strain gauges) was tested under a static condition. The calibration of the displacement transducers (plastic film potentiometer) was compared for static and dynamic recordings. Force measurement was found precise (deviation < 1.1%) after alterations in the calibration procedure. Displacement measurement was affected by punch tilting and the application of the transducers. If tilting of punches was not considered, the deviation of displacement measurement from the true value (using steel tablets as a reference) was found up to 110 μm. By modifying the original PPA system by supplementing additional displacement transducers in the adjacent turret positions of the punches and adding a custom electronic device (Tilting Compensation Device), the accuracy of distance measurement was improved to 18.1 μm (±3.64). Furthermore, machine and tooling deformation were recorded and found different under static and dynamic conditions. Correction of punch displacement for elastic deformation therefore should preferably be made from dynamic recordings.