Determination of the quenching correction factors for plastic scintillation detectors in therapeutic high-energy proton beams
- 6 November 2012
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 57 (23), 7767-7781
- https://doi.org/10.1088/0031-9155/57/23/7767
Abstract
Plastic scintillation detectors (PSDs) have many advantages over other detectors in small field dosimetry due to their high spatial resolution, excellent water equivalence and instantaneous readout. However, in proton beams, the PSDs undergo a quenching effect which makes the signal level reduced significantly when the detector is close to the Bragg peak where the linear energy transfer (LET) for protons is very high. This study measures the quenching correction factor (QCF) for a PSD in clinical passive-scattering proton beams and investigates the feasibility of using PSDs in depth-dose measurements in proton beams. A polystyrene-based PSD (BCF-12, ϕ0.5 mm × 4 mm) was used to measure the depth-dose curves in a water phantom for monoenergetic unmodulated proton beams of nominal energies 100, 180 and 250 MeV. A Markus plane-parallel ion chamber was also used to get the dose distributions for the same proton beams. From these results, the QCF as a function of depth was derived for these proton beams. Next, the LET depth distributions for these proton beams were calculated by using the MCNPX Monte Carlo code, based on the experimentally validated nozzle models for these passive-scattering proton beams. Then the relationship between the QCF and the proton LET could be derived as an empirical formula. Finally, the obtained empirical formula was applied to the PSD measurements to get the corrected depth-dose curves and they were compared to the ion chamber measurements. A linear relationship between the QCF and LET, i.e. Birks' formula, was obtained for the proton beams studied. The result is in agreement with the literature. The PSD measurements after the quenching corrections agree with ion chamber measurements within 5%. PSDs are good dosimeters for proton beam measurement if the quenching effect is corrected appropriately.Keywords
This publication has 20 references indexed in Scilit:
- Dosimetric performance and array assessment of plastic scintillation detectors for stereotactic radiosurgery quality assuranceMedical Physics, 2011
- Determination of average LET of therapeutic proton beams using Al2O3:C optically stimulated luminescence (OSL) detectorsPhysics in Medicine & Biology, 2010
- Characterizing the response of miniature scintillation detectors when irradiated with proton beamsPhysics in Medicine & Biology, 2008
- Monte Carlo investigation of collimator scatter of proton-therapy beams produced using the passive scattering methodPhysics in Medicine & Biology, 2007
- Current developments in proton therapy: a reviewPhysics in Medicine & Biology, 1993
- Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: I. Physical characteristics and theoretical considerationsPhysics in Medicine & Biology, 1992
- Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: II. Properties and measurementsPhysics in Medicine & Biology, 1992
- Cerenkov light generated in optical fibres and other light pipes irradiated by electron beamsPhysics in Medicine & Biology, 1992
- Analysis of response data for several organic scintillatorsNuclear Instruments and Methods, 1970
- The non-linear response of the plastic scintillator NE102Nuclear Instruments and Methods, 1967